ECG
10-leadmode:LeadsR/RL,F/LL,C1/V1,andC5/V5
PacemakerdetectionisdoneonleadsI,II,andVsimultaneously.
TheVleadplaysacriticalroleinpacemakerdetection.
Selectingthepacemakerdetection
Pacemakerdetectionmustbeturnedon.However,youmaydisablepacemakerevent
processingbyturningoffpacemakerdetection.Whenpacemakerdetectionisturned
off,themonitoringdeviceignorespacemakerpulsedetections,whichmayadversely
affecttheheartrateaccuracyofthemonitoringdevice.
1.SelecttheHRparameterwindow.
2.SelectAdvanced>Page1.
3.SelectavaluefromthePacemakerDetectionlist.
Sensitive(Pace2):Includespacemakermarkerpulsesinthewaveformwitha
moresensitivelowerthresholdonthepacepulseamplitude.
Normal(Pace1):Detectsanddrawspacemakermarkerpulsesinthe
waveform.Thedetectionislesssensitivetoelectromagneticinterference.This
optionisrecommendedwhentheinterferencelevelishigh,forexampledue
toanLVADdeviceorinfusionpumps.
Off:Pacemakerpulsearenotdetectedormarkedwithpacemakermarker
bars.Pacemakerpulsesareshownastheyappearinthewaveformsignal.
Pacemakerdetectiontroubleshooting
ProblemSolution
Howdoesactivatingpacemakerdetectionimpact
monitoring?
Beatsthatwouldotherwisebeclassiedas
ventricularareinsteadclassiedasV-pacedifa
ventricularpacemakereventisdetected.
Residualpacemakerenergythatmightotherwise
appearintheECGisremoved,andapacemaker
enhancedspikeisplacedintheECG.
OntheECGwaveform,pacemakerdetection
isindicatedbyuniform,uprightpacemaker
enhancementspikesintheECGdata.
Howcanpacemakerdetectionbeimproved?Possibleproblemsinclude:
Heartratedoublecounting.
Inaccuratealarmsforlowheartrateorasystole.
Pacemakerspikesnotrecognizedbythe
software.
FalsePVCdetectionsandarrhythmiaalarms.
Possiblesolutionsinclude:
RelearnECG.
Re-preparethepatientskin,replacethe
electrodes,andadjusttheelectrodeplacement.
Tryanalternateelectrodeplacement.
Trysingle-leadanalysis,ifavailable.
114CARESCAPEONE2094480-001