ECG
■10-leadmode:LeadsR/RL,F/LL,C1/V1,andC5/V5
●PacemakerdetectionisdoneonleadsI,II,andVsimultaneously.
●TheVleadplaysacriticalroleinpacemakerdetection.
Selectingthepacemakerdetection
Pacemakerdetectionmustbeturnedon.However,youmaydisablepacemakerevent
processingbyturningoffpacemakerdetection.Whenpacemakerdetectionisturned
off,themonitoringdeviceignorespacemakerpulsedetections,whichmayadversely
affecttheheartrateaccuracyofthemonitoringdevice.
1.SelecttheHRparameterwindow.
2.SelectAdvanced>Page1.
3.SelectavaluefromthePacemakerDetectionlist.
●Sensitive(Pace2):Includespacemakermarkerpulsesinthewaveformwitha
moresensitivelowerthresholdonthepacepulseamplitude.
●Normal(Pace1):Detectsanddrawspacemakermarkerpulsesinthe
waveform.Thedetectionislesssensitivetoelectromagneticinterference.This
optionisrecommendedwhentheinterferencelevelishigh,forexampledue
toanLVADdeviceorinfusionpumps.
●Off:Pacemakerpulsearenotdetectedormarkedwithpacemakermarker
bars.Pacemakerpulsesareshownastheyappearinthewaveformsignal.
Pacemakerdetectiontroubleshooting
ProblemSolution
Howdoesactivatingpacemakerdetectionimpact
monitoring?
●Beatsthatwouldotherwisebeclassiedas
ventricularareinsteadclassiedasV-pacedifa
ventricularpacemakereventisdetected.
●Residualpacemakerenergythatmightotherwise
appearintheECGisremoved,andapacemaker
enhancedspikeisplacedintheECG.
●OntheECGwaveform,pacemakerdetection
isindicatedbyuniform,uprightpacemaker
enhancementspikesintheECGdata.
Howcanpacemakerdetectionbeimproved?●Possibleproblemsinclude:
■Heartratedoublecounting.
■Inaccuratealarmsforlowheartrateorasystole.
■Pacemakerspikesnotrecognizedbythe
software.
■FalsePVCdetectionsandarrhythmiaalarms.
●Possiblesolutionsinclude:
■RelearnECG.
■Re-preparethepatientskin,replacethe
electrodes,andadjusttheelectrodeplacement.
■Tryanalternateelectrodeplacement.
■Trysingle-leadanalysis,ifavailable.
114CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017