Parameterspecications
NIBPtotalcycletime20to40secondstypical(dependingonheartrate,pressure,and
motionartifact)
Automaticcuffdeationconditions●Poweroff
●Adultandchildcuffcycletimeexceeding125seconds
●Infantcuffcycletimeexceeding90seconds
●Adultandchildcuffpressureexceeding300mmHg(40.0kPa)
●Infantcuffpressureexceeding150mmHg(20.0kPa)
Invasivepressurestandardscompliance
ThesystemwithCARESCAPEPressurecomplieswithIEC60601-2-34:2011-05.
TheCARESCAPEPressuresensorinterfacecableandcompatibletransducersareTYPE
CFDEFIBRILLATION-PROOFAPPLIEDPARTsperIEC60601-2-34Clause201.6.
Invasivepressureperformancespecications
Numberofinvasivepressurechannels
2
Invasivepressuremeasurementrange
-98to349mmHg(-13.1to46.5kPa)
Invasivepressuremeasurement
accuracy
±0.5%±1.50mmHg(excludingtransducer)
±4%or±4mmHg,whicheverisgreater(includingtransducer)
Invasivepressuredisplayresolution
1mmHg
InvasivepressureunitsofmeasuremmHgorkPa
Invasivepressurezerobalancerange±150mmHg(±20.0kPa)
Invasivepressurezerobalanceaccuracy±1mmHg(±0.1kPa)
Invasivepressurepulseraterange0to360bpm
Invasivepressurepulserateaccuracy±2%or±2bpm,whicheverisgreater
Invasivepressurepulseratedisplay
resolution
1bpm
Invasivepressuretransducersites,site
name,anddisplayedvalues
●Arterial(Art):Systolic,diastolic,mean,rate
●Femoral(Fem):Systolic,diastolic,mean,rate
●Femoralvein(FemV):Mean
●Pulmonaryartery(PA):Systolic,diastolic,mean
●Centralvenouspressure(CVP):Mean
●Intra-cranialpressure(ICP):Mean
●Leftatrial(LAP):Mean
●Pressurechannel(P1,P2):Mean
●Rightatrial(RAP):Mean
●Rightvein(RVP):Mean
●Umbilicalartery(UAC):Systolic,diastolic,mean,rate
●Umbilicalvein(UVC):Mean
2094480-001CARESCAPEONE319
CARESCAPE ONE User Manual DRAFT 26 October 2017