CO₂
SelectingCO
2
average
YoucanselectatimewindowthatisusedtoselectthedisplayedCO
2
value.The
highestpeakEtCO
2
valuewithinthetimewindowisselectedastheCO
2
value
displayed.
1.SelecttheCO
2
parameterwindow.
2.SelectSetup.
3.SelectanoptionfromtheCO2Averagelist.
SelectingtheFiO
2
level
ThepresenceofalargeconcentrationofoxygencausestheCO
2
leveltoappearlower
thantheactualvalue.UsethisoptiontocompensateforthepresenceofO
2
.
1.SelecttheCO
2
parameterwindow.
2.SelectSetup.
3.SelectanoptionfromtheFiO2levellist:21-40or40-100percent.
SelectingtheN
2
Olevel
ThepresenceofN
2
OcausestheCO
2
valuetoappearhigherthantheactualvalue.
UsethisoptiontocompensateforthepresenceofN
2
O.
1.SelecttheCO
2
parameterwindow.
2.SelectSetup.
3.SelectanoptionfromtheN2Olevellist:0-40or40-80percent.
Selectingtheapneaalarmlimit
Theapneaalarmcanbeenabledtoselectatimedelaybeforeanapneaalarmis
generated.ThedelayisdenedwiththeApnealimitsecondssetting.Ifanything
otherthanthedefault(20seconds)isselected,thenumberofsecondsselectedforthe
delaydisplaysintheparameterwindow.
1.SelecttheCO
2
parameterwindow.
2.SelectAlarms>Page2.
3.SelectanoptionfromtheApnealimitsecondslist.
ZeroingtheLoFloSidestreamModule
WhenthemessageCO2zerorequiredappearsinthealarmmessagesarea,orthe
messageZerorequiredappearsintheCO
2
parameterwindow,theLoFloSidestream
Moduleneedstobezeroed.
NOTE
Forbestresults,allowveminutesformodulewarm-up
beforestartingthezero.
YoushouldalsozerotheLoFloSidestreamModuleafterconnectingorreconnecting
asampleline.
1.DisconnectthepatientfromCO
2
monitoring.
2.ConnecttheLoFloSidestreamModuleand,ifnecessary,waituntilthesensor
warm-upmessagedisappears.