F
Parameterspecications
Parameterspecications
WARNING
INACCURATERESULTS.Donotuseorstoretheequipment
outsidethespeciedtemperature,humidity,oraltitude
ranges,oroutsidethespeciedperformancerange.Using
orstoringtheequipmentoutsidethespeciedoperating
environmentoroutsidethespeciedperformancerangemay
causeinaccurateresults.
ECGstandardscompliance
ThesystemwithCARESCAPEECGcomplieswithIEC60601-2-27:2011-03.
NOTE
Moderateandmaximumreducedbandwidthsdonotcomply
withallrequirementsoftheIEC60601-2-27standard.
TheCARESCAPEECGenclosureandinterfacecablearetreatedasTYPECF
DEFIBRILLATION-PROOFAPPLIEDPARTsperIEC60601-2-27Clause201.6.2.
CompatibleleadwiresareTYPECFDEFIBRILLATION-PROOFAPPLIEDPARTs.
ECGperformancespecications
ECGheartraterange20to300bpm
NOTE
Valuesintherangeof301to360bpmare
displayedas300bpm.
ECGheartrateaccuracy±1%or±1bpm,whicheverisgreater
ECGheartrateresolution1bpm
ECGheartratesensitivity0.5mVpeak
ECGheartrateresponsetimeIndicatesanewheartrateforastepincreaseof80to120bpm
andastepdecreaseof80to40bpmin<10s
Heartrateaccuracyandresponseto
irregularrhythm
HeartratecalculationoperateswithirregularrhythmsofIEC
60601-2-27Clause201.7.9.2.9.101b)4),accordingtoFigure
201.101,asfollows:
Ventricularbigeminy:80bpm
Slowalternatingventricularbigeminy:59bpm
Rapidalternatingventricularbigeminy:126bpm
Bidirectionalsystoles:110bpm
2094480-001CARESCAPEONE297
CARESCAPE ONE User Manual DRAFT 26 October 2017