Invasivepressures
1.Selecttheinvasivepressureparameterwindow.
2.SelectSetup>Page2.
3.SelecttheheartratesourcefromthePrimaryHRSourcelist.
NOTE
Onlychannelsthathavebeenconguredwitharterial
pressurelabelswillappearinthelist.
Showingthepulserateintheinvasivepressureparameter
window
NOTE
ThissettingisavailableforArtX,FemX,orUACXinvasive
pressurechannelsonly.
1.Selecttheinvasivepressureparameterwindow.
2.SelectSetup>Page2.
3.SelectShowPulseRate.
ShowingtheCPPvalueintheICPparameterwindow
Avalidmeanarterialpressureisrequiredtocomputethecerebralperfusionpressure
(CPP)value.
1.Selecttheinvasivepressureparameterwindow.
2.SelectSetup>Page2.
3.SelectShowCPP.
UsingtheIPchannelstandby
Ifyouwishtoprepareandzeroachannelbeforehand,youcanusethechannel
standby.
1.Selecttheinvasivepressureparameterwindow.
2.SelectSetup>Page1.
3.SelectStandby[channellabelchannelnumber].Thechannellabelandchannel
numbertextdependsonthechannellabelandchannelnumberoftheinvasive
pressureparameterwindowthatwasselected,forexample,StandbyArt2.
ChannelalarmsandmeasurementaredisableduntilActivate[channellabel
channelnumber]isselectedorpressureisdetected.Thechannellabeland
channelnumbertextdependsonthechannellabelandchannelnumberofthe
invasivepressureparameterwindowthatwasselected,forexample,Activate
Art2.
Selectingtheinvasivepressurenoisereductionlter
NOTE
Ifarterialpressureisusedtotriggertheintra-aorticballoon
pump,usethe40Hzpressurelter.
1.Selecttheinvasivepressureparameterwindow.
2.SelecttheAdvancedtab.
170CARESCAPEONE2094480-001