Alarms
Powerloss
duration
Alarmsettingstatusafterapowerloss
Upto2hoursThealarmsettingsthatareineffectbeforethepowerlossarerestored
automatically.
Greaterthan2
hours
Thealarmsettingsrevertbacktotheuserdefaultsettings(startup
prole).Youmustrecongureanypatient-specicalarmsettings.
AlarmdatastoredinClinicallogs
AccesstotheClinicallogsisaservice-levelfunctionanditispasswordprotected.
CARESCAPEONEstoresarecordofpatient-relatedalarmsandinformationmessages
aswellasanyadjustmentstothealarmlimitsintheClinicallogs.
Storedalarmdataduringapowercycleorpowerloss
IftheCARESCAPEONEgoesthroughapowercycleoralossofpower,thestored
alarmdataintheClinicallogisnotaffected.Thealarmdataremainsstoredinthe
Clinicalloguntilthemonitorautomaticallyclearstheoldeststoreddatatoallownew
datatobestored.