ECG
SettingtheSVTlength
ThissettingdetermineshowmanyconsecutiveSVCsareneededtotriggertheSV
Tachyalarm.
1.SelectAlarmsSetup.
2.SelecttheArrhythmiatab.
3.SelectAtrialAlarms>SVTachytab.
4.SelectavaluefromtheSVTLengthlist.
SettingHRforSVT
ThissettingdeterminestheminimumvaluefortheHRtotriggertheSVTachyalarm.
1.SelectAlarmsSetup.
2.SelecttheArrhythmiatab.
3.SelectAtrialAlarms>SVTachytab.
4.SelectavaluefromtheHRforSVT/minlist.
Arrhythmiaalarmmessages
NOTE
Aclinicianmustanalyzethearrhythmiainformationin
conjunctionwiththeotherclinicalndings.
AlarmmessageArrhythmiaanalysisAlarmprioritydefaultArrhythmiadetection
criteria
AFibFull
Allsoftwarepackages
exceptNICU
Accordingtopriority
setting
AbsenceofP-wavesand
irregularRR-interval.
Accel.Ventric.FullAccordingtopriority
setting
Acceleratedventricular
rhythm-RunofPVCs
witharunlengthofat
leastsixbeatsandthe
raterequirementshave
notbeenmetfortheHR
forVTach/minsetting.
AsystoleFull,LethalHighHRdecreasedtozero,or
beatdetectionhasnot
occurredinthelast5
seconds.
Bigeminy
FullAccordingtopriority
setting
EveryotherbeatisPVC
(N-V-N-V-N-V).
CoupletFullAccordingtopriority
setting
TwoconsecutivePVCs
aredetectedbetween
normalbeats,N-V-V-N.
Thecouplinginterval
betweenthePVCsmust
belessthan600ms.
IrregularFull
NICUsoftwarepackage
only
Accordingtopriority
setting
Sixconsecutivenormal
RRintervalsvaryby100
msormore.
118CARESCAPEONE2094480-001