Safety
Electricalwarnings
WARNING
POWERSUPPLY.Alwaysconnectthedevicepowercabletoa
properlyinstalledpoweroutletwithprotectiveearthcontacts
beforeconnectinganyotherinterfacecables.Iftheintegrity
oftheprotectiveearthconductorisindoubt,disconnectthe
monitorfromthepowerlineanduseitwiththebatteryoption.
Iftheinstallationdoesnotprovideforaprotectiveearth
conductor,disconnectthedevicepowercablefromthepower
lineafterhavingdisconnectedallotherinterfacecables.All
devicesofasystemmustbeconnectedtothesamepower
supplycircuit.Deviceswhicharenotconnectedtothesame
circuitmustbeelectricallyisolatedwhenoperated.
WARNING
EXCESSIVELEAKAGECURRENT.Donotuseamultiplesocket
outletorextensioncordinanMEsystem.
WARNING
EXCESSIVELEAKAGECURRENT.Toavoidsummationofleakage
currentswheninterfacingthedevicewithotherequipment,
thedevicesmayonlybeinterconnectedwitheachother
ortopartsofthesystemwhenithasbeendeterminedby
qualiedbiomedicalpersonnelthatthereisnodangertothe
patient,theoperator,ortheenvironmentasaresult.Inthose
instanceswherethereisanyelementofdoubtconcerning
thesafetyoftheconnecteddevices,theusermustcontact
themanufacturersconcerned(orotherinformedexperts)for
properuse.Inallcases,safeandproperoperationshould
beveriedwiththeapplicablemanufacturer’sinstructions
foruse,andsystemstandardsIEC60601-1clause16must
becompliedwith.
WARNING
EXCESSIVELEAKAGECURRENT.ConnectonlycertiedUL
60950/IEC60950equipmenttotheCARESCAPEDockF0RJ-45
connection.
WARNING
INTERFACINGOTHEREQUIPMENT.Connectonlyitemsthatare
speciedaspartofthesystemandascompatible.Formore
information,seethesupplementalinformationprovided.
WARNING
INTERFACINGOTHEREQUIPMENT.Devicesmayonlybe
interconnectedwitheachotherortopartsofthesystemwhen
ithasbeendeterminedbyqualiedbiomedicalpersonnel
thatthereisnodangertothepatient,theoperator,orthe
environmentasaresult.Inthoseinstanceswherethereis
anyelementofdoubtconcerningthesafetyoftheconnected
devices,theusermustcontactthemanufacturersconcerned
(orotherinformedexperts)forproperuse.Inallcases,safe
andproperoperationshouldbeveriedwiththeapplicable
manufacturer’sinstructionsforuse,andsystemstandardsIEC
60601-1clause16mustbecompliedwith.
WARNING
Donotunderanycircumstancesremovethegrounding
conductorfromthepowerplug.Alwayscheckthatpower
cordandplugareintactandundamaged.
WARNING
Duringintracardiacapplicationofadevice,adebrillatorand
pacemakerwhoseproperfunctioninghasbeenveriedmust
bekeptathand.