SpO₂
SpO
2
canbethePrimaryHRSource.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelecttheheartratesourcefromthePrimaryHRSourcelist.
ShowingtheSpO
2
pulserate
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelectShowPulseRate.
AdjustingtheSpO
2
pulsebeepvolume
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page2.
3.AdjustthevolumewiththeBeatVolumeselector.
Selectingthebeatsource
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page2.
3.SelectthebeatsourcefromtheBeatSourcelist.
Thebeatsourceindicator
willappearbesidethechosenbeatsourceonthescreen,
andthebeatsoundwillreectthebeatofthatsource.
Settingthevariablebeattone
TheSpO
2
variablebeattonecanbesettoaffectthepitchoftheheartratebeeptone.
Whenturnedon,theheartratebeeppitchchangesaccordingtoincreasingand
decreasingSpO
2
values.HigherSpO
2
saturationvalueshaveahigherpitchbeep
tone,andlowervalueshavealowerpitch.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page2.
3.SelectVariableBeatToneSource>SpO2orOff.
Formoreinformation,seethesupplementalinformationprovided.
MasimoSETdataaveragingandupdating
ForMasimoSETtechnology,whenusingthedefaultaveragingtimeof8seconds,
thereisamaximumdata-averagingsignalprocessingtimeof10secondsfromreal
timeplusanadditionaldelayof2secondstoupdatethedisplayedwaveform.
SelectingtheSpO
2
averagingtime
NOTE
MasimoSETtechnologyandMasimosensorsorGETruSignal
technologyandTruSignalsensorsonly.
YoucanhaveanaverageoftheSpO
2
measurementonscreen,andyoucanselect
howmanysecondsareusedforthisaveraging.WithMasimoSETtechnology,the
2094480-001CARESCAPEONE145