Defaultsettings
Alarmdelaysettings
SelectAlarmDelaystodenethealarmdelaysettings.Thefollowingtableliststhefactory
defaultsettings.
Setting
ED
ICUNICUORPACU
PR(SpO2/IP)high/low
5s
SpO2high
5s
SpO2low
5s
RR(Imped.)high
15s
(PediatricandInfant
proles:10s)
10s15s
(PediatricandInfant
proles:10s)
RR(Imped.)low
15s
(PediatricandInfant
proles:10s)
10s15s
(PediatricandInfant
proles:10s)
Invasivepressuresalarmssettings
SelectProleSettings>InvasivePressures>Alarmstodenetheinvasivepressures
alarmssettings.
SelectArterialtabtoadjusttheinvasivepressuresarterialalarmsprolesettings.The
followingtableliststhefactorydefaultsettings.
Setting
ED
ICUNICUORPACU
ArtSysHighLimit/Low
Limit
180/80
(Neuroprole:160/90;
Adolescentprole:150/80;
Pediatricprole:130/70;
Infantprole:120/50)
(USonlyInfantprole:100/50)
90/40
(Cardiac,
Full-term,and
Infantproles:
100/40)
180/80
(Neuroprole:
160/90;
Perfusion
prole:
120/40;
Adolescent
prole:
150/80;
Pediatric
prole:
130/70;
Infantprole:
120/50)
(USonly
Infantprole:
100/50)
180/80
(Neuroprole:
160/90;
Adolescent
prole:
150/80;
Pediatric
prole:
130/70;
Infantprole:
120/50)
(USonly
Infantprole:
100/50)
ArtSysAlarmOnEnabled