SpO₂
SpO
2
waveformstability
ThestabilityofthedisplayedSpO
2
valuescanalsobeusedasanindicationofsignal
validity.ToaidyouinsuccessfulSpO
2
monitoring,messagesareprovidedintheSpO
2
parameterwindow.
SpO
2
wavelengthsandopticaloutputpower
GETruSignal,Nellcor
sensorswithOxiMax
technology,andMasimoSETpulse
oximetrytechnologiesarecalibratedtodisplayfunctionalsaturation.
Thisinformationmaybeusefultoclinicianssuchasthoseperformingphotodynamic
therapy:
NellcorsensorswithOxiMaxtechnologycontainLEDsthatemitredlightata
wavelengthofapproximately660nmandinfraredlightatawavelengthof
approximately900nm.ThetotalopticaloutputpowerofthesensorLEDsisless
than15mW.
MasimoSETpulseoximetrysensorscontainLEDsthatemitredlightata
wavelengthofapproximately660nmandinfraredlightatawavelengthof
approximately905nmforRDSET,M-LNCS,LNCS,andLNOP,andapproximately
653nmand880nmforRDSET,M-LNCS,LNCS,andLNOPtipclips.Thetotaloptical
outputpoweroftheLEDsislessthanorequalto15mW.
GETruSignalpulseoximetrysensorscontainLEDsthatemitredlightata
wavelengthofapproximately663nmandinfraredlightatawavelengthof
approximately890or940nm.ThemaximumopticaloutputpowerforeachLEDis
lessthan15mW.
SpO
2
measurementandinterference
ThesetypesofinterferencecaninuencethefunctionofSpO
2
:
Incorrectsensorapplication,e.g.,sensorplacementonanextremitywithablood
pressurecuff,arterialcatheter,orintravascularline,sensorappliedtootightly.
Intravasculardyes,suchasidocyanineormethyleneblue.
Externallyappliedcoloringagentswithopaquematerialsinhighambientlight
conditions,e.g.,conditionscreatedfromoneormoreofthefollowingsources:
Surgicallights,especiallyxenonlightsources
Bilirubinlamps
Fluorescentlights
Infraredheatinglamps
Directsunlight
Excessivepatientactivity
Venouspulsation
Dysfunctionalhemoglobin
Poor(low)peripheralperfusion
Arterialocclusionproximaltothesensor
Lossofpulse(cardiacarrest)
Electromagneticinterference(EMI)
Ventilator-inducedpressurechange
150CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017