EasyManua.ls Logo

GE CARESCAPE ONE - Pulse Rate Accuracy of Masimo SET Technology with Masimo Sensors; Pulse Rate Accuracy of GE Trusignal Technology with Trusignal Sensors; Spo 2 Performance Specifications

GE CARESCAPE ONE
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameterspecications
NeonatespecicationsareshownforOxiMaxMAXNsensorswiththepulseoximeter.
Clinicalfunctionalityhasbeendemonstratedonapopulationofhospitalizedneonate
patients.TheobservedSpO
2
accuracywas2.5%inastudyof42patientswithagesof
1to23days,weightfrom750to4,100grams,and61observationsmadespanning
arangeof77%to98%SaO
2
.
Specicationappliestomonitoringcableperformance.Readingaccuracyinthe
presenceoflowperfusion(detectedIRpulsemodulationamplitude0.03%-1.5%)was
validatedusingsignalssuppliedbyapatientsimulator.SpO
2
andpulseratevalues
werevariedacrossthemonitoringrangeoverarangeofweaksignalconditionsand
comparedtotheknowntruesaturationandpulserateoftheinputsignals.
Nellcor™sensorsmotionperformancewasvalidatedduringacontrolledhypoxia
bloodstudyoveranSaO
2
spanof70%to98.9%andaconvenience-sampleheart
rangeof41-105bpm.Subjectsperformedrubbingandtappingmovements1-2cmin
amplitudewithaperiodicintervals(randomlychanging)witharandomvariationin
frequencybetween1-4Hz.Theaveragepercentrangewasvalidatedusingsynthetic
signalsfromapatientsimulatorthatcomprisedrepresentativecardiacandsignal
artifactcomponents.Applicability:OxiMaxMAXA,MAXAL,MAXP,MAXI,andMAXN
sensors.
PulserateaccuracyofMasimoSETtechnologywithMasimo
sensors
MasimoSETTechnologywithMasimosensorshasbeenvalidatedforpulserate
accuracyfortherangeof25-240bpminbenchtoptestingagainstaBiotekIndex
2simulator.Thevariationequalsplusorminusonestandarddeviation,which
encompasses68%ofthepopulation.
PulserateaccuracyofGETruSignaltechnologywith
TruSignalsensors
GETruSignaltechnologywithTruSignalsensorshasbeenvalidatedforpulserate
accuracyoverthespeciedrangeinbenchtoptestingagainstapatientsimulator.
SpO
2
performancespecications
NOTE
TechnicalspecicationsgivenforCARESCAPESpO
2
Masimoare
basedonMasimoCorporationownedMasimoSETtechnology
andsensors.Formoredetailedinformation,refertothesensor
instructionsforuse.
NOTE
TechnicalspecicationsgivenforCARESCAPESpO
2
Nellcorare
basedonCovidienAGownedtechnologyandsensors.Formore
detailedinformation,refertothesensorinstructionsforuse.
ThefollowingspecicationsapplytoallCARESCAPESpO
2
technologiesunlessotherwise
indicated.
Pulseoximetrysaturationdisplay
resolution
1digit(%ofSpO
2
)
Pulseoximetryperipheralpulserate
displayresolution
1bpm
304CARESCAPEONE2094480-001