EasyManua.ls Logo

GE CARESCAPE ONE - Alarm Priorities and Escalation Times for Impedance Respiration

GE CARESCAPE ONE
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarmspecications
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
Trigeminy
alarm,
wavef.
0s,accordingtoprioritysetting:high,medium,low,off,
info
VBradyalarm,
wavef.
0s,accordingtoprioritysetting:high,medium,low,off,
info
VFib/VTachalarm,
param.,
wavef.
0s
VTachalarm,
param.,
wavef.
0s,accordingtoprioritysetting:high,medium,low,info
Thealarmpriorityisalwayshighifthedurationisover
30s,thesetHRhighlimitisexceeded,andtheHRis
over180inNICUswpackageorover150inallother
swpackages.
VT>2
alarm,
wavef.
0s,accordingtoprioritysetting:high,medium,low,off,
info
Alarmprioritiesandescalationtimesforimpedance
respiration
Thefollowingtableshowsthemessageswiththeirprioritiesandescalationtimes.Inthe
table,themessagelocationisindicatedbythefollowing:alarm=alarmarea;param.=
parameterwindow;andwavef.=waveformarea.Inthefollowingtable,0s=nointended
delay.Formoreinformationonalarmmessages,refertotheuserinformationprovided.
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
Apnea(Imped.)ORswpackage,
adjustablelimit
seconds
alarm,
wavef.
Xseconds
(X=
adjustable
limit)
X+40sX+102s
Apnea(Imped.)
PACU,ED,ICU,
NICUsw
packages,
adjustablelimit
seconds
alarm,
wavef.
Xseconds
(X=
adjustable
limit)
X+20sX+40s
Apnea(Imped.)Alarmpriority
settoinfo,low,
mediumorhigh
alarmXs,accordingtoprioritysetting:high,medium,low,
info(X=adjustablelimit)
Apnea
deactivated
param.
0s
Artifact
param.,
wavef.
0s
Cardiacartifactalarm
0s
278CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Related product manuals