Impedancerespiration
Impedancerespirationcompatibilitylimitations
FordetailedinformationregardingCARESCAPEONE,CARESCAPEParameter,and
accessorycompatibility,seethesupplementalinformationprovided.
Respirationsafetyprecautions
Respirationwarnings
WARNING
Makesurethattheleadwiresetclipsorsnapsdonottouch
anyelectricallyconductivematerialincludingearth.
WARNING
Theimpedancerespirationmeasurementisinherently
verysensitiveasitmeasuresverysmallphysiological
signals(changesofimpedanceofthepatient’schest
area).Electromagneticinterferencemaycauseerroneous
measurementsatvariousfrequencies,forexample
interferencewiththesignal/waveform,leadingtorespiration
ratereadingsinconsistentwiththepatient'struerespiration
rate.Ifyounoticethis,useanotherformofrespiration
monitoring,forinstanceend-tidalCO
2
.
WARNING
Whenusinganelectrosurgeryunit,notethatthe
measurementcablesdonotincorporatemeanstoprotect
againstburnsincaseofadefectiveESUreturnelectrode.To
avoidburnsatthemonitormeasurementsites,ensurethe
following:
PropercontactoftheESUreturnelectrodetothepatient.
ESUreturnelectrodeneartheoperatingarea.
Measurementelectrodes,leadwiresandprobesfarfrom
thesurgicalsiteandtheESUreturnelectrode.
WARNING
APNEAEVENTS.Thedevicemaynotdetectallepisodesof
inadequatebreathing,nordoesitdistinguishbetweencentral,
obstructive,andmixedapneaevents.
WARNING
DEFIBRILLATORPRECAUTIONS.Patientsignalinputslabeled
withtheCFandBFsymbolswithpaddlesareprotectedagainst
damageresultingfromdebrillationvoltages.Toensure
properdebrillatorprotection,useonlytherecommended
cablesandleadwires.