ECG
DisplayingtheECGgrid
YoucanhaveareferencegridintheECG1,ECG2,andECG3waveformareas.Thegrid
pointswillbeat200mshorizontallyand0.5mVvertically.
1.SelecttheHRparameterwindow.
2.SelectAdvanced>Page2.
3.SelecttheECGGridcheckboxtodisplaythegrid.
ECGalarmlimits
TheHRAlarmscanbesettoSingleorMultiplethroughMonitorSetup>Defaults&
Service>DefaultSetup>CareUnitSettings>Parameters.Thissettingispassword
protected.
TheSingleheartratealarmsettingallowsyoutosetonecommonHRlimitformultiple
sources(e.g.,ECG,SpO
2
,Art)andthePVCandSVCalarmlimitsforECGfromtheHR
AlarmsArrhythmiatab.Withthissettingactivated,turningofftheSpO
2
orinvasive
pressure(Art,Fem,UAC)HRalarmlimitsalsoturnsofftheprimaryHRalarm,and
adjustingtheSpO
2
orinvasivepressure(Art,Fem,UAC)HRlimitvaluesalsoadjuststhe
primaryHRlimitvalue.
TheMultipleheartratealarmsettingallowsyoutosetaprimaryheartrate/pulserate
sourceanduptothreeindividualheartrate/pulseratealarmsandlimitsfromtheHR
AlarmsArrhythmiatab.ItalsoallowsyoutosetPVCandSVCalarmlimitsforECG.
WiththeMultipleheartratesettingactivated,turningofftheSpO
2
HRalarmlimits
doesnotturnofftheprimaryHRalarm.Theinvasivepressurechannelsthatappear
intheHRAlarmsArrhythmiatabdependonthechannelsthatareconguredfor
arterialpressures.Amaximumof2channelswillappearifbothinvasivepressure
channelsareconguredtoarterialpressures,(e.g.,Fem1,Art2).
TheHRalarmdelayiscongurablefrom0to20secondsinincrementsof5seconds
fortheinvasivepressureandSpO
2
parameters.Formoreinformation,seethe
supplementalinformationprovided.
SettingHRalarmlimitsforasingleHRsource
1.SelecttheHRparameterwindow.
2.SelecttheHRAlarmsArrhythmiatab,thenselecttheHRtab.
3.Checkthattherequiredalarmisturnedon.
4.SelectAlarmOntoenablethealarms.
5.Adjustthealarmlimitswiththearrows.
SettingHR/PRalarmlimitsformultipleHRsources
1.SelecttheHRparameterwindow.
2.SelectHRAlarmsArrhythmia>HR/PR.
3.Checkthattherequiredalarmisturnedon.
4.SelecttheAlarmOncheckboxforthosealarmsyouwishtoset.
5.Adjustthealarmlimitswiththearrows.
110CARESCAPEONE2094480-001