Invasivepressures
Invasivepressurepointstonote
Thisequipmentissuitableforuseinthepresenceofelectrosurgery,astested
accordingtoIEC60601-2-49clause202.6.2.101Electrosurgeryinterference.
Forinformationonmaterialsusedinaccessoriesandtheirbiocompatibility,refer
totheinstructionsforuseintheaccessorypackage.
Debrillatordischargemayaffecttheinvasivepressuremeasurement.Refertothe
supplementalinformationprovidedfordetailsregardingtherecoverytimefrom
debrillatordischarge.
DonotturnontheIABPalgorithmunlessaballoonpumpisinuse.
Invasivepressuremeasurementsetup
Invasivepressureequipmenttopatientconnectionwith
CARESCAPEONE
1.CARESCAPEONE
2.Fluidbagwithpressureinfusor
3.Transducersetup
4.CARESCAPEPressure
Connectingtheinvasivepressuretransducerandcable
1.Preparethetransducerkitaccordingtothemanufacturer’sinstructions.
2.ConnectthetransducersetuptotheCARESCAPEPressure.
3.Removeentrappedairfromwithinthetransducersetupbytappingthesetup
andbyturningitintodifferentpositions.
4.ConnecttheCARESCAPEPressuretotheCARESCAPEONE.
5.Connectthetransducertothepatientline.
Checkingtheinvasivepressuremeasurement
1.Makesurethatallthetransducersarezeroedcorrectly.
166CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017