Defaultsettings
Setting
ED
ICUNICUORPACU
P1-P2MeanHigh
Limit/LowLimit
15/0
(Infantprole:10/0)
5/015/0
(Infantprole:10/0;Perfusion
prole:15/-5)
P1-P2MeanAlarmOnEnabledEnabled
(Neuroand
Trauma
proles:
Disabled)
Enabled
SelectVentricular&Atrialtabtoadjusttheinvasivepressuresventricularandatrialalarm
prolesettings.Thefollowingtableliststhefactorydefaultsettings.
Setting
ED
ICUNICUORPACU
RAPMeanHigh
Limit/LowLimit
15/0
(PediatricandInfant
proles:10/0)
(USonly:15/-5)
(USonly:ICUsoftware
packagePediatricandInfant
proles:15/0)
5/015/0
(Pediatricand
Infantproles:
10/0)
15/0
(USonly:
15/-5)
(Pediatricand
Infantproles:
10/0)
RAPMeanAlarmOnEnabled
RVPMeanHigh
Limit/LowLimit
35/5
(PediatricandInfant
proles:25/5)
25/535/5
(PediatricandInfant
proles:25/5)
RVPMeanAlarmOnEnabled
LAPMeanHigh
Limit/LowLimit
20/5
(PediatricandInfant
proles:15/5)
10/220/5
(PediatricandInfant
proles:15/5)
LAPMeanAlarmOnEnabled
Invasivepressuressettingsforlabels
Selectinganewchannellabelforachannelwillapplythefollowinglabeldefaultsettings
tothecurrentpatient.Onceanewlabelisselected,thechannel’sdefaultsettingscanbe
changedforthecurrentpatientandalsosavedasthechannel’sprolesettingdefaults.
Whenanychannellabelotherthanthedefaultchannellabelisselected:
FilterHz:12
WhenArtisselectedasthechannellabel:
CatheterDisconnect:Enabled
DisplayFormat:Sys/Dia(Mean)
ShowPulseRate:Disabled
PrimaryArterialSource:AUTO
SmartBP:Enabled
ScalemmHg(waveform):
2094480-001CARESCAPEONE
337
CARESCAPE ONE User Manual DRAFT 26 October 2017