J
Electromagneticcompatibility
IEC60601-1-2
ThesystemcomplieswithIEC60601-1-2:2007-03.
Accordingtoparameter-specicIEC60601-2-xseriesstandardrequirementsforESU
(electrosurgicalunit)tests,theequipmentisprotectedagainstmalfunctioncaused
byelectrosurgery.
EssentialperformanceinEMC
Parameter
Essentialperformance
GeneralNolossofdisplayorusersettings.
Nolossofpatientdata,modeofoperation,orstoreddataduringlossofACmains
orbatterypower>10s.
Noinvalidorerroneousalarms.
Abilitytogenerateatechnicalalarmduringabnormaloperation.
ECG
Nofalsebeatdetectionthatproducesanincorrectheartratemeasurementgreater
than±10%or±5bpm,whicheverisgreater.
Nowaveformamplitudeerror>20%.
Impedance
respiration
Noerrorinrespirationrategreaterthan±1breathperminute.
Invasiveblood
pressure
Noinvasivepressureerrorgreaterthan±4%ofreadingor±0.5kPa4mmHg),
whicheverisgreater.
Temperature
Noerrorsintemperaturereadings>0.3°C.
CO
2
(RGM)
NoerrorinCO
2
greaterthan±(volumefractionof+8%ofGasLevel)withthegas
levelsetbetween71-100mmHg..
Nochangeinrespirationgreaterthan±1breathperminute.
CO
2
owrate<60ml/min.
SpO
2
:GETruSignal,
Nellcor,Masimo
Noerrorsinpulserategreaterthan±2bpmforGETruSignalSpO
2
orgreaterthan
±3bpmforNellcorandMasimoSpO
2
.
Noerrorinsaturationgreaterthan±4%.
Non-invasiveblood
pressure
Noerrorforthemeasurementofthecuffpressuregreaterthan±2mmHg0.3kPa).
Debrillatorsync
markeroutput
TimedelayfrompeakofECGR-wavetosynchronizationpulse35ms.
2094480-001CARESCAPEONE387
CARESCAPE ONE User Manual DRAFT 26 October 2017