ECG
STdetection
AbouttheSTanalysis
Ifenabled,STanalysisstartsautomaticallyaftertheECGleadshavebeenconnected
andQRSdetectionhasstarted.Oncetheprogramhascompletedthelearningphase,
STvaluesareupdatedevery10seconds.
STdetection
Duringthelearningperiod,thealgorithmusestheisoelectricreferenceandtheJ+
referencepointstocalculatetheSTvalues.Thealgorithmautomaticallysearches
fortheJandISOpoints.
STdetectionmeasurementlimitations
STvaluesmaybeaffectedbysuchfactorsassomedrugsormetabolicand
conductiondisturbances.
ChangesinheartratemayaffectST.
TheSTalgorithmhasbeentestedforaccuracyoftheSTsegmentdata.The
signicanceoftheSTsegmentchangesneedstobedeterminedbyaphysician.
STdetectionpointstonote
STsegmentdeviationsarenotdisplayedforpatientswithventricularpacemakersor
iftherhythmisconsideredasfromventricularorigin.
StartingtheSTdetection
1.SelecttheSTparameterwindoworselecttheHRparameterwindowandtheST
tab.
2.SelectOnfromtheSTAnalysislist.
STalarmlimits
YoumaysettheSTalarmlimitsforindividualleads.
Settingalarmlimitsforindividualleads
1.SelecttheSTparameterwindow.
2.SelectSTAlarms.
3.SelectAlarmOnforanECGleadtoadjustitsalarmlimits.
Ifthealarmislocked,thereisalocksymbolbesidethecheckboxandthe
selectionisnotavailable.
4.SetHighandLowalarmlimits.