CO₂
CO
2
troubleshooting
ProblemSolution
CO
2
waveformseemsabnormal.Checkthepatient:
Checkforanychangeinpatientcondition.
Ifthepatienthasacannula,checkforproper
placementofthenasalprongs.
Ifthepatientisbreathingthroughthemouth,
ensurethatanoral/nasalaccessoryisused.
Abnormalwaveformscanbeseenifthepatient
istalking.
Checkthesamplelinesandairwayadapters.
Replaceifnecessary.
Elevatedbaseline.Checkthepatientstatus.
ChecktheLoFloSidestreamModule.
MakesurethepatientisnotbreathingCO
2
.Check
thattheamountoffreshgasissufcientforthe
patient,andcheckthattheCO
2
absorberisin
goodcondition.
Makesurethesamplecellisproperlyconnected(a
clickishearduponconnection)beforezeroing.
Iftheelevatedbaselineisseenwhentheaccessory
issamplingroomair,performazero.
Excessivesecretionsinsamplelines.Checktheorientationoftheairwayadapterport.
Useaccessorieswithdehumidicationtubing.
IfthemessageSamplelineblockedappears,
replacethesamplelinesandairwayadapters.
Measurementisnotworking.ChecktheLoFloSidestreamModuleandadapter
positioning.
CARESCAPECO
2
orLoFloSidestreamModuleisnot
working.
Contactqualiedservicepersonnel.
Readingsseeminaccurate.Checkthepatientstatus.
Checkthatthegasexhaustlineisnotkinkedor
blocked.
Checkthebreathingcircuitforpossibleleaks.
2094480-001CARESCAPEONE195