Messages
Messagesrelatedtoimpedancerespiration
measurementwithCARESCAPEONE
Forinformationregardingalarmprioritiesandescalationtimes,seethesupplemental
informationprovided.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
●al.area=alarmarea
●param.=parameterwindow
●wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
●Apnea(Imped.)
●APN
●Apnea
●al.area
●param.
●wavef.
Nobreathingdetected.●Checkthepatientstatus.
●Checktheventilatorand
breathingstatus.
●Apneadeactivated
●param.
Thecasehasjustbeen
started/patientadmittedonthe
monitor,orthemeasurement
hasjustbeenstarted.
●Wait.Themessage
disappearsafterthe
respirationrateis≥3/min.
●Artifact
●CardiacArtifact
●param.,
wavef.
●al.area
Cardiacartifacthasbeen
detected.
●Checkthepatientstatus.
●Selectalternateleadsto
monitor.
●Increasesensitivitysetting.
●Selectalternatelead
placement.
●Relearnrespiration.
●Checkdevice
●Imped.malfunction
●param.,
wavef.
●al.area
Hardwareorsoftwarefaultin
CARESCAPEECG:
●Cablefault
●Congurationfailure
●Noupdateddata
●Datasampledrift
●ReconnecttheCARESCAPE
ECG.
●ReplacetheCARESCAPEECG.
●Iftheproblempersists,
contactqualiedservice
personnel.
●LeadIfailed
●param.,
wavef.
LeadIhasbeenselectedfor
measurement,buttheRAorLA
electrodeisofforisnotadhered
tothepatientwellenoughtoget
asufcientsignalforrespiration
monitoring.
●Checktheelectrodesand
theirconnections.
●Youmayneedtopreptheskin
andreplacetheelectrode.
●LeadIIfailed
●param.,
wavef.
LeadIIhasbeenselectedfor
measurement,buttheRAorLL
electrodeisofforisnotadhered
tothepatientwellenoughtoget
asufcientsignalforrespiration
monitoring.
●Checktheelectrodesand
theirconnections.
●Youmayneedtopreptheskin
andreplacetheelectrode.
2094480-001CARESCAPEONE213