ECG
SettingPVCalarmlimits
Availableonlywhenfullarrhythmiadetectionisenabled.
1.SelecttheHRparameterwindow.
2.SelectHRAlarmsArrhythmia>PVC/SVCArrhythmia.
3.CheckthatthePVCalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntoenablethealarms.
5.Adjustthealarmlimitswiththearrows.
SettingSVCalarmlimits
Availableonlywhenfullarrhythmiaisenabled.
1.SelecttheHRparameterwindow.
2.SelectHRAlarmsArrhythmia>PVC/SVCArrhythmia.
3.CheckthattheSVCalarmisturnedon.
Ifafeatureisnotactive,thealarmlimitsaregreyedout.
4.SelectAlarmOntoenablethealarms.
5.Adjustthealarmlimitswiththearrows.
ECGalarmpriorities
YoucansetthealarmprioritiesforcertainECGalarmsthroughAlarmsSetup>Alarm
Priorities>ECG.TheallowedprioritiesaredenedintheCareUnitSettings,and
theyarepasswordprotected.
ECGmeasurementpracticalities
Alternatepulseratesource
Thealternatepulseratesourceallowsclinicianstoacquireapulseratefromasource
otherthanECG(Art,Fem,UAC,orSpO
2
).Thefollowingcircumstancesmaywarrant
theuseofanalternatepulseratesource:
Excessiveartifactduetoanelectricalinterferencefromequipment(e.g.,
electrosurgicaldevice).
Excessivepatientmovementcausingsignicantartifact(e.g.,seizureactivity).
Inabilitytousestandardleadplacement(e.g.,burns).
IntelliRatealgorithm
TheIntelliRatealgorithmextractsinformationfrommultiplephysiologicalsignals
(ECG,SpO
2
,andArt,FemorUACwhenselectedastheprimaryarterialchannel)
andappliesrule-basedlogictodeterminewhichheartratesourcehasthehighest
likelihoodofbeingaccurate.TheIntelliRatealgorithmcanalsosuppressalarmsfor
Asystole,VFib/VTach,andVTachifthealgorithmdeterminesthatthereisahigh
likelihoodofafalsearrhythmiacall.Byreportingthemostaccuraterate,thetrended
pulserateismoreaccurate,andoccurrencesoffalsepulseratelimitviolationalarms
2094480-001CARESCAPEONE111