Alarmstandardscompliance
ThesystemcomplieswithIEC60601-1-8:2012–11.
NOTE
Legacyalarmtonesdonotmeettheaudiorequirements
denedinIEC60601-1-8Clause6.3.3.1.
Auditoryalarmvolume,IECtones
TestedinaccordancewithIEC60601-1-8subclause6.3.3.2withalarmvolumecontrolset
tomaximum.
CARESCAPEONE
AlarmvolumesettingMaximumsoundpressurelevel
6
Level6correspondstothe
factorydefaultminimum
allowedsettingforvolume
level.
Highpriorityalarm:53dBA
Mediumpriorityalarm:50dBA
Lowpriorityalarm:49dBA
10
Highpriorityalarm:76dBA
Mediumpriorityalarm:75dBA
Lowpriorityalarm:73dBA
Auditoryalarmvolume,legacytones
TestedinaccordancewithIEC60601-1-8subclause6.3.3.2withalarmvolumecontrolset
tomaximum.
2094480-001CARESCAPEONE261