Alarmspecications
AlarmLimitrangeLimitincrementAlarmdelay
RR(CO2)high4to60breaths
perminute
NICUsw
package:4to
100breathsper
minute
1breathper
minute
21s(average)(20to22srange)
RR(CO2)low4to60breaths
perminute
NICUsw
package:4to
100breathsper
minute
1breathper
minute
21s(average)(20to22srange)
Technicalalarmdelayspecications
Thefollowingtableliststhealarmdelaysfortechnicalalarms.Anyalarmdelaysthatareless
than5saregivenas<5s.
AlarmParameterAlarmdelay
Analogoutputmalfunction
<5s
Arrhythmiapaused
ECG
21s(average)(20to22srange)
ArtXdisconnect
FemXdisconnect
UACXdisconnect
Disconnected
Invasivepressures
8s
Artifact
ECG
<5s
Calibrationfail
T1Calibrationfail
T2Calibrationfail
Temperature<5s
Callservice
NIBP
22s(average)(22srange)
CheckCO2adapter
CO
2
42s(average)(41to42srange)
Checkdevice
CO2devicemalfunction
CO
2
<5s
Checkdevice
ECGmalfunction
ECG
<5s
Checkdevice
Imped.malfunction
Impedancerespiration
<5s
2094480-001CARESCAPEONE269
CARESCAPE ONE User Manual DRAFT 26 October 2017