ECG
aregreatlyreduced.Thealternatepulseratesourcevaluereplacesthestandardheart
ratevalueintheHRparameterwindow.
Autoalgorithm
TheAUTOalgorithmselectstherstavailableheartratesourcebasedona
pre-denedparameterpriority:
1.ECG
2.Primaryarterialsite
3.SpO
2
ECGtroubleshooting
ProblemSolution
ECGsignalisnoisyornoQRSisdetectedEnsurethatthepatientisnotshivering.
Checktheelectrodequalityandpositioning.Do
notplaceelectrodesonbodyhair,bonescloseto
skin,layersoffatandmajormuscles.Pre-gelled
electrodesarerecommended.
ChangetheleadinECG1tothebestavailable
signalandconsiderusingtheSingleleadmode.
ConsiderusingSize2xor4x.
TryanalternativelocationfortheValeadto
improvesignalquality.Insomecases,likeifthe
patienthasasignicantheartfailure,changingfor
examplefromV5toV1canresultinaconsiderable
differenceinthesignalamplitude.
Selecttheappropriateltertoapplytothe
displayedwaveformsbyselectingtheHR
parameterwindow>Advanced>Page1>
WaveformFilter.ThisdoesnotaffectQRS
detectionorarrhythmiaanalysis.
Checkallcableconnectors.
WhyaretheQRScomplexesmarkedwithanX?Debrillatorsynchronizationmarkersareidentied
withanXontheQRScomplexes.
Numericvaluesarereplacedbydashes.ReconnecttheCARESCAPEECG.
ReplacetheCARESCAPEECG.
Iftheproblempersists,contactqualiedservice
personnel.
Aboutthe12RL
feature
NOTE
12RLisnotavailablewiththeNICUsoftwarepackage.
NOTE
Althoughreconstructed(interpolated)leadscanbeviewed
ontheCARESCAPEONEinstandalonemode,theCARESCAPE
ONEmustbeconnectedtoahostmonitorasanacquisition
devicetoobtaina12RLreport.
112CARESCAPEONE2094480-001