SpO₂
WhenusingtheMaximumSensitivitysetting,performanceoftheSensorOff
detectionmaybecompromised.Iftheinstrumentisinthissettingandthesensor
becomesdislodgedfromthepatient,thepotentialforfalsereadingsmayoccur
duetoenvironmental“noise”suchaslight,vibration,andexcessiveairmovement.
Donotloopthepatientcablingintoatightcoilorwraparoundthedevice,asthis
candamagethepatientcabling.
AdditionalinformationspecictotheMasimosensorscompatiblewiththepulse
oximeter,includinginformationaboutparameter/measurementperformance
duringmotionandlowperfusion,maybefoundinthesensor’sdirectionsforuse
(DFU).
CablesandsensorsareprovidedwithX-Caltechnologytominimizetheriskof
inaccuratereadingsandunanticipatedlossofpatientmonitoring.Refertothecable
orsensordirectionsforuseforthespecieddurationofthepatientmonitoringtime.
SpO₂measurementguidelines
GETruSignaltechnologyandsensormeasurement
guidelines
ThefollowingmeasurementguidelinesapplytoGETruSignalSpO
2
technology:
Thetimeperiodforacquiringameasurementaverageisadjustable.
TheSpO
2
waveformcorrespondsto(butisnotproportionalto)thearterialpressure
waveform.
OnlyTruSignalsensorsaresupported.
UsethefollowingguidelineswhenusingTruSignalsensorsandcables:
ReadthesensorinstructionsforuseoftheSpO
2
sensorbeforeusingit.
Periodicallyinspectextensioncablesandsensorsfordamage.
Donotusedamagedsensors.
RefertothecleaninginstructionsintheinstructionsforuseofreusableTruSignal
sensors.
DonotuseNIBPorconstrictinginstrumentsonthesameappendageasthe
SpO
2
sensor.
MasimoSETtechnologyandsensormeasurement
guidelines
Withmotion,theplethysmographicwaveform(orSpO
2
waveform)isoften
distortedandmaybeobscuredbytheartifact.WithMasimoSETtechnology,the
plethysmographicwaveformisnotanindicationofsignalqualityorvalidity.Evenwith
awaveformobscuredbyartifact,MasimoSETtechnologyisabletoreadthroughthe
noiseandlocatethearterialpulsation.
AlthoughMasimoSETtechnologyprocessesSpO
2
measurementsdifferentlythan
otherSpO
2
technologies,thefunctionandappearanceisessentiallythesameasother
technologies.ThefollowingmeasurementguidelinesapplytoMasimoSETtechnology
only:
Thetimeperiodforacquiringameasurementaverageisadjustable.
2094480-001CARESCAPEONE141