Defaultsettings
CO
2
defaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>CO2>Setuptoadjust
thesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
Scale%SelecttheCO
2
scale.
0-6
CO2Sweep
Speed
SelecttheCO
2
sweepspeed.
6.25mm/s
CO2Average
Selectthetime
windowthatis
usedtoselect
thedisplayed
CO
2
value.
10s
FiO2level%Selectthe
levelofFiO
2
usedtomake
compensation
tomeasured
CO
2
values.
21-40
N2Olevel%Selectthe
levelofN
2
O
usedtomake
compensation
tomeasured
CO
2
values.
0-40
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>CO2>Alarms>Page
1orPage2toadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhe
factorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
EtCO
2
alarm
limits
Selectthe
high/lowalarm
limits(%).
8/3
(NeuroandTrauma
proles:6.0/2.7.)
8/38/3
(Neuro
prole:
6.0/2.7;
Perfusion
prole:8/0)
8/3
(Neuro
prole:
6/2.7)
FiCO
2
alarm
limits
Selectthe
high/lowalarm
limits(%).
3/OFF
FiCO
2
alarms
on/off
Enablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
364CARESCAPEONE2094480-001