Defaultsettings
Softwareproles
Eachsoftwarepackagesupportsuptoeightactiveproles.Aproleisagroupof
uniquesettingssuitedtoaparticularcareunitorpatientdemographicwithinthe
broadersoftwarepackage(e.g.,alarmlimits,screenlayouts,trendsandsnapshot
settings).Prolesettingscanbecustomized.Thefollowingtableliststhefactory
defaultprolesforeachsoftwarepackage.Thedefaultproleislistedrst.
ED
ICUNICUORPACU
MedicalMedical
Premature
GeneralGeneral
Trauma
SurgicalCardiacPerfusion
Invasive
CardiacCardiacFull-term
Invasive
Regional
NeuroNeuro
InfantRegional
Neuro
Respiratory
Trauma
Neuro
PreRegional
AdolescentAdolescentAdolescentAdolescent
PediatricPediatricPediatricPediatric
InfantInfant
Therearefour
pre-dened
prolesfor
NICUandfour
undened
proles(Prole
5toProle8)
thatcanbe
customizedas
needed.
InfantInfant
Softwaresettings
NOTE
TheenabledsoftwareproleandCARESCAPEParameters
useddeterminewhichcongurablesettingsdisplayon
theCARESCAPEONE.Somesettingsmaynotappearinall
congurations.
NOTE
Somesettingsvarybyconguration.Thesesettingsare
indicatedwhereapplicable.
Ifthemonitorresetswhilemonitoringapatient,careshouldbetakentoconrmthe
settingsareappropriateforthecurrentpatient,makingchangeswherenecessary.
Whenthesupplymainstotheequipmentisinterruptedformorethan30seconds
whilethemonitorison,themonitorshutsdown.Whenrestartedwithin2hours,itwill
continuemonitoringwiththepreviouslyselectedusersettings.
Careunitsettings
Careunitsettingsarepasswordprotected.Careunitsettingsareestablishedduring
installation/congurationofthemonitor.Instructionstoaccessthecareunitsettings
areprovidedinthisdocument.Thecareunitsettingsfactorydefaultvaluesarealso
provided.
Careunitsettingsapplytothesoftwarepackageandallassociatedproles.Care
unitsettingshavetwovalues:
●Currentvalues:Valuesdisplayedonthemonitor,savedduring
conguration/installationofthemonitortomeettheneedsofaparticularclinical
environment.