18
Messages
MessagesrelatedtoECGmeasurement
Forinformationregardingalarmprioritiesandescalationtimes,seethesupplemental
informationprovided.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
AFibal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Accel.Ventric.al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Arrhoff
param.
Thearrhythmiadetectionlevel
issettoOff.
Ifarrhythmiadetectionis
needed,setthedetection
leveltoFullorLethal.
Arrhythmiapaused
Arrhpaused
al.area
wavef.
ECGchannelshavenotbeen
availableforanalysisforthelast
20secondsduetoexcessive
artifact,ortheinternalHR
calculationhasnotbeen
updatedforthelast30seconds,
orinsingleleadanalysiswhen
arrhythmiadetectionissetto
FullorLethal,theECG1lead
hasnotbeenactiveforthelast5
seconds.
Checkthepatientstatus.
Checkelectrodeplacement.
Preparethepatient’sskinat
electrodesites.
Changeormoveelectrodes.
2094480-001CARESCAPEONE209
CARESCAPE ONE User Manual DRAFT 26 October 2017