Messages
Message
Location
PossiblecausesSuggestedactions
Artifactwavef.Muscleartifact,highorlow
frequencynoise,orpacemaker
artifact.
Checkelectrodecontact.
Checkleadplacement.
Performskinpreparation.
Reposition/replace
electrodes.
Requestthepatienttoremain
still.
Checkifanyelectricaldevices
aretooclosetothepatient.
Asystoleal.area,
param.,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Bigeminy
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Brady
HRLOW
HR(ECG)low
al.areaPhysiologicalalarm.
Measurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Checkdevice
ECGmalfunction
wavef.
al.area
Hardwareorsoftwarefaultin
CARESCAPEECG:
Cablefault
Noupdateddata
Datasampledrift
Congurationfailure
ReconnecttheCARESCAPE
ECG.
ReplacetheCARESCAPEECG.
Iftheproblempersists,
contactqualiedservice
personnel.
Coupletal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Deviceoverheating
MoveECGdevicetoa
coolerlocation
wavef.
al.area
CARESCAPEECGsurface
temperatureistoohigh.
ChecktheCARESCAPEECG
placementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
ReplacetheCARESCAPEECG.
ECGdevicefailure:
Callservice
al.area
CARESCAPEECGstartupor
initializationfailure.
ReconnecttheCARESCAPE
ECG.
ReplacetheCARESCAPEECG.
Iftheproblempersists,
contactqualiedservice
personnel.
ECGdevice
overheated.Shutting
down
al.areaCARESCAPEONEturnsoffthe
CARESCAPEECGbecauseits
surfacetemperatureistoo
high.
ChecktheCARESCAPEECG
placementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
ReplacetheCARESCAPEECG.
210CARESCAPEONE2094480-001