Messages
Message
Location
PossiblecausesSuggestedactions
●Artifact●wavef.Muscleartifact,highorlow
frequencynoise,orpacemaker
artifact.
●Checkelectrodecontact.
●Checkleadplacement.
●Performskinpreparation.
●Reposition/replace
electrodes.
●Requestthepatienttoremain
still.
●Checkifanyelectricaldevices
aretooclosetothepatient.
●Asystole●al.area,
param.,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●Bigeminy
●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●Brady
●HRLOW
●HR(ECG)low
●al.areaPhysiologicalalarm.
Measurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Checkdevice
●ECGmalfunction
●wavef.
●al.area
Hardwareorsoftwarefaultin
CARESCAPEECG:
●Cablefault
●Noupdateddata
●Datasampledrift
●Congurationfailure
●ReconnecttheCARESCAPE
ECG.
●ReplacetheCARESCAPEECG.
●Iftheproblempersists,
contactqualiedservice
personnel.
●Couplet●al.area,
wavef.
Physiologicalalarm.●Checkthepatientstatus.
●Deviceoverheating
●MoveECGdevicetoa
coolerlocation
●wavef.
●al.area
CARESCAPEECGsurface
temperatureistoohigh.
●ChecktheCARESCAPEECG
placementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
●ReplacetheCARESCAPEECG.
●ECGdevicefailure:
Callservice
●al.area
●CARESCAPEECGstartupor
initializationfailure.
●ReconnecttheCARESCAPE
ECG.
●ReplacetheCARESCAPEECG.
●Iftheproblempersists,
contactqualiedservice
personnel.
●ECGdevice
overheated.Shutting
down
●al.area●CARESCAPEONEturnsoffthe
CARESCAPEECGbecauseits
surfacetemperatureistoo
high.
●ChecktheCARESCAPEECG
placementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
●ReplacetheCARESCAPEECG.
210CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017