CO₂
5.Connectthesamplelinetothepatient’sbreathingcircuitifanairwayadapter
accessoryisused,orplaceonthepatientifanasalcannulaisused.
Aboutsidestreamkits
CAUTION
Thenasalandoral/nasalcannulakitsareintendedforsingle
use.Donotreuseorsterilizethecannulakitassystem
performancewillbecompromised.
SidestreamkitsareintendedformonitoringCO
2
ofnon-intubatedpatients.
●Selectthekitthathasthebesttandmostcomfortforthepatient.Forexample,
thekitlabeledasPediatricmayhavethebesttandbemorecomfortableon
asmall-staturedadultthantheadultcannulakit.Thecannulakitsareequally
functionalregardlessofwhichisused.
●Verifythatthecannulakitisclean,dryandundamaged.Replacethecannulakit
ifnecessary.
●Insertthesamplecellintothesamplecellreceptacle.Youwillhearaclickwhenthe
samplecellisproperlyinserted.
●ZerotheLoFloSidestreamModule.
●Placethenasalcannulakitontothepatient:
Seetheaccessory-specicinstructionsforuseformoredetailedinformation.
UsingtheCO₂measurement
PumponoroffwithCARESCAPECO
2
ThepumpisautomaticallyturnedonwhenasamplelineisinsertedintotheLoFlo
SidestreamModule.Whenconnectedproperly,youwillhearaclick.Whenthesample
lineisremoved,thepumpturnsoffimmediately.Whenablockedlineoccurs,the
pumpturnsoffafter2minutesofunsuccessfullytryingtoremovetheblockage.
SelectingtheCO
2
scale
IfEtCO
2
isabove6%(45mmHg),changethescaleforthecapnogram.
1.SelecttheCO
2
parameterwindow.
2.SelectSetup.
3.SelectanoptionfromtheScalelist.
SelectingtheCO
2
sweepspeed
Thisselectionchangesthespeedatwhichthecapnogramisdisplayed.