Non-invasivebloodpressure
2.SelectNIBPAuto>StartCycling.
3.StopthemeasurementbyselectingNIBPAuto>StopCycling.
SettingthecycletimebetweenNIBPmeasurements
ToautomaticallymeasureNIBPatsettimeintervals,youmustrstsetthecycletime.
1.SelecttheNIBPparameterwindow>Page1.
2.SelectthecycletimefromtheCycleTimelist.
AutomaticNIBPmeasurementsandmonitorclock
synchronization
Clocksynchronizationtiming(cyclingsync)automaticallysynchronizesyourautomatic
NIBPmeasurementtimeintervalswiththemonitorclock.Forexample,ifautomatic
measurementsareinitiatedforveminuteintervalsat4:02,therstmeasurementis
takenimmediatelyat4:02.Thenextmeasurementwillbetakenat4:05(intervaland
clockarenowsynchronized).Allmeasurementswillcontinueatveminuteintervals
(i.e.,4:10,4:15,etc.)
Therewillalwaysbeatleasta30seconddelaybetweentwoconsecutiveNIBP
measurementsduringautocycling.Ifanautomaticmeasurementcompleteswith
lessthan30secondstothenextscheduledmeasurement,themonitorwilldelaythe
scheduledmeasurementuntil30secondshavepassed.Thecyclingsynchronization
isnotdoneduringthese30secondsbutitwillbedoneafterthedelayedauto
measurementstarts.
Exampleswitha5minutecyclingtime:
Completionwithlessthan30secondsto
thenextscheduledmeasurement
Completionafterthenextscheduled
measurementshouldhavestarted
●Firstautomeasurementstarts:4:59:00
●Firstautomeasurementcompletes:
4:59:40
●Secondautomeasurementstarts:5:00:20
(notclocksynchronized)
●Thirdautomeasurementstarts:5:05:00
(clocksynchronized)
●Firstautomeasurementstarts:4:59:00
●Firstautomeasurementcompletes:
5:00:10
●Secondautomeasurementstarts:5:05:00
STATmode
NOTE
NotavailableintheNICUsoftwarepackage.
TheSTATmodeinitiatesacontinuouscycleofmeasurementsforveminutes.The
messageSTATdisplaysintheNIBPparameterwindowwhenSTATisstarted.Thereis
alsoagraphicalbarindicatorintheparameterwindow.AnewNIBPmeasurement
startsafterthepreviousmeasurementcompletes.Theamountoftimebetween
measurementsvaries.Itisatleastfoursecondsforadultandchildcuffsizesand
atleasteightsecondsforinfantcuffsize.Theearlysystolicvalueismeasured
anddisplayeduntilthenalresultisavailable,butisneverproducedfortherst
measurementinaseriesofSTATmodemeasurements.Afterveminutes,themonitor
automaticallyreturnstothepreviouslyselectedcyclingintervalormanualmode.
158CARESCAPEONE2094480-001