EasyManua.ls Logo

GE CARESCAPE ONE - ECG Cautions; ECG Measurement Limitations; ECG Points to Note; ECG Measurement Setup

GE CARESCAPE ONE
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ECG
WARNING
HEARTRATEALARMINTERFERENCE.Poorcablepositioning
orimproperelectrodepreparationmaycauselineisolation
monitortransientstoresembleactualcardiacwaveformsand
thusinhibitheartratealarms.Tominimizethisproblem,follow
properelectrodeplacementandcablepositioningguidelines
providedwiththisdevice.
ECGcautions
CAUTION
Thepatient'sskinmaybecomeirritatedafterprolonged
contactwithelectrodegeloradhesive.
ECGmeasurementlimitations
ThemonitorwilldisplayaLeadsoffmessageinaninputoverloadcondition,or
upondisconnectionofelectrodeleadwires.
ECGpointstonote
Thisequipmentissuitableforuseinthepresenceofelectrosurgery,astested
accordingtoIEC60601-2-49clause202.6.2.101Electrosurgeryinterference.
Forinformationonmaterialsusedinaccessoriesandtheirbiocompatibility,refer
totheinstructionsforuseintheaccessorypackage.
DebrillatordischargemayaffecttheECGmeasurement.Refertothesupplemental
informationprovidedfordetailsregardingtherecoverytimefromdebrillator
discharge.
Pre-gelledECGelectrodesarerecommended.Checktheexpirationdate.
Makesuretheelectrodegelismoist.
Makesuretheelectrodeshavegoodskincontact.
Replaceallelectrodesatleastevery24to48hours.
SelecttheDeactivateECGLeadsOffoptiontoremoveaLeadsoffmessagefrom
thedisplaywhenacableisoff.
ECGmeasurementsetup
ECGequipmenttopatientconnectionwithCARESCAPEONE
2094480-001CARESCAPEONE99