Non-invasivebloodpressure
NIBPtroubleshooting
ProblemSolution
NIBPmeasurementdoesnotworkor
thevaluesseemunstable.
Checkthatthecufftubingisnotbent,stretched,compressed,
orloose.
Checkthecuffpositionandcufftubeconnection.
Preventmotionartifact.
UseNIBPcuffsofcorrectsize.
NIBPmeasurementdoesnotstart.Ensurethatthecuffsizehasbeenselected.
Whyisthecuffre-inating
automatically?
Thecufftargetpressuremustbehigherthanthepatient’ssystolic
pressuretoobtainanaccuratesystolicanddiastolicmeasurement.
Ifasystolicbloodpressurecannotbefound,asystolicreadingis
searchedbyre-inatingthecufftoahigherpressure.Duringa
systolicsearch,themaximumcuffinationpressurewillnotexceed
thenormalpressurerangeofthecuff.Formoreinformation,refer
tothetechnicalspecications.
Acontrolmeasurementmaybetakingplace.IfthemeasuredNIBP
valueexceedsthealarmlimits,asinglelowpriorityalarmsounds
andanewmeasurementisautomaticallytaken.Ifthenewvalue
(thecontrolmeasurement)alsoexceedsthealarmlimitsthealarm
priorityescalatestomedium.InManualmodeandSTATmode
thereareatleastfoursecondsbetweentherstmeasurementand
thecontrolmeasurementforAdultandChildcuffs,eightseconds
forInfantcuffs.InAutomodethereareatleast30seconds
betweentherstmeasurementandthecontrolmeasurement.
UnexpectedNIBPreadingsEnsurethatthecorrectcuffsizeforthepatienthasbeen
selected.
Checkthatthecuffispositionedcorrectly.
Reducesourcesofmotionartifact.
Calibratethedevice.
164CARESCAPEONE2094480-001