Electromagneticcompatibility
Parameter
Essentialperformance
ECGanalogoutput●NoerrorsinECGwaveformamplitudegreaterthan±10%.
Invasiveblood
pressureanalog
output
●Noerrorsininvasivebloodpressureanalogoutputgreaterthan1V/100mmHg±
1.5%±1.0mmHgor±0.5%±2mmHg,whicheverisgreater.
Electromagneticcompatibilitysafety
precautions
Electromagneticcompatibilitywarnings
WARNING
Theuseofaccessories,transducersandcablesother
thanthosespeciedmayresultinincreasedemissionsor
decreasedimmunityperformanceoftheequipmentor
system.
WARNING
Otherequipmentmayinterferewiththesystem,evenifthat
otherequipmentcomplieswithCISPRemissionrequirements.
WARNING
Thedevice/systemshouldnotbeusedadjacentto,orstacked
with,otherequipment.Consultqualiedpersonnelregarding
device/systemconguration.
Electromagneticcompatibilitycautions
CAUTION
UseofknownRFsources,suchascell/portablephones,
orotherradiofrequency(RF)emittingequipmentnearthe
systemmaycauseunexpectedoradverseoperationof
thisdevice/system.Consultqualiedpersonnelregarding
device/systemconguration.
CAUTION
Changesormodicationstothisdevice/systemnotexpressly
approvedbyGEmaycauseEMCissueswiththisorother
equipment.
Electromagneticemissions
Thedeviceisintendedforuseintheelectromagneticenvironmentspeciedbelow.
Thecustomerortheuserofthedeviceshouldassurethatitisusedinsuchan
environment.