Parameterspecications
MeasuredA
RMS
values
Range
A
RMS
90-100%1.45%
80-90%1.22%
70-80%1.41%
Overallclaimedaccuracyvalue
Range
A
RMS
70-100%2%
TCI
MeasuredA
RMS
values
Range
A
RMS
90-100%1.05%
80-90%1.67%
70-80%2.43%
Overallclaimedaccuracyvalue
Range
A
RMS
70-100%3.5%
AdditionalaccuracyinformationforGETruSignalsensors
ThetableinformationprovidessupplementaldataanalysisforGETruSignalsensors’
measurementaccuracy.
ThefollowingmodiedBland-AltmanplotsshowSpO
2
databysensortype.
2094480-001CARESCAPEONE313
CARESCAPE ONE User Manual DRAFT 26 October 2017