Defaultsettings
SettingDescription
ED
ICUNICUORPACU
Selectthehigh/low
adultalarmlimits.
180/80
(Adolescentprole:150/80.ORNeuroprole:160/80.)
Selectthehigh/low
childalarmlimits.
130/70120/70130/70
Systolicalarm
limits
Selectthehigh/low
infantalarmlimits.
120/5090/40
(Cardiac,
Full-term
andInfant
proles:
100/40.)
120/50
Systolicalarm
on
Enablealarmwhen
limitsareviolated.
Enabled
Selectthehigh/low
adultalarmlimits.
140/60
(Adolescentprole:
110/50)
140/60140/60
(Adolescentprole:
110/50)
Selectthehigh/low
childalarmlimits.
100/5090/50100/50
Meanalarm
limits
Selectthehigh/low
infantalarmlimits.
80/4060/30
(Cardiac,
Full-term
andInfant
proles:
70/40.)
80/40
MeanalarmonEnablealarmwhen
limitsareviolated.
Disabled
(Infant
prole:
Enabled)
Disabled
(Adoles-
centpro-
le:En-
abled
EnabledDisabled
(Infantprole:Enabled)
Selectthehigh/low
adultalarmlimits.
100/40
(Adolescentprole:80/40;ORsoftwarepackage
Neuroprole:90/40.)
Selectthehigh/low
childalarmlimits.
80/4060/4080/40
Diastolicalarm
limits
Selectthehigh/low
infantalarmlimits.
70/3050/20
(Cardiac,
Full-term
andInfant
proles:
60/20.)
70/30
Diastolicalarm
on
Enablealarmwhen
limitsareviolated.
Disabled
Invasivepressuresdefaultsettings
Invasivepressurechannellabelsandchannelnumbersbothhavedefaultsettings.If
youchangeapressurechannellabel,thechannel’sdefaultsettingswillchangetothe
selectedlabel’sdefaultsettings.Onceanewlabelisselected,thechannel’sdefault
2094480-001CARESCAPEONE359