Defaultsettings
settingscanbechangedforthecurrentpatientandalsosavedasthechannel’sprole
settingdefaults.Forexample,ifthechannel1labelArt1ischangedtoRAP1,theRAP
1labeldefaultsettingsareappliedtochannel1forthecurrentpatient.Thecurrent
patientsettingscanbechangedandalsosavedthechannel1prolesettingsdefaults.
Ifnolabelisselectedforachannel,thenthechannel’sdefaultsettingsareused.
Invasivepressurechannel1defaultstothechannellabelArtintheED,ICU,OR,and
PACUsoftwarepackages.ItdefaultstothechannellabelUACintheNICUsoftware
package.Invasivepressurechannel2defaultstothechannellabelCVPintheED,ICU,
OR,andPACUsoftwarepackages.ItdefaultstothechannellabelUVCintheNICU
softwarepackage.
Art1pressuresitedefaultsettings
ThedefaultP1pressuresitefortheED,ICU,OR,andPACUsoftwarepackageisArt.Select
MonitorSetup>MainSetup>ParameterSetup>Page1>Art1>Setup>Page1or
Page2toadjustthesesettingsforthecurrentpatient.Thefollowingtableliststhefactory
defaultsettings.
SettingDescription
ED
ICUORPACU
LabelSelectthepressuresite
label.
Art
ScalemmHgSelectthewaveform
scale.
0-200
(Adolescentprole:0-170;
Pediatricprole:0-140;
Infantprole:0-120)
0-200
(Perfusion
prole:
0-120;
Adolescent
prole:
0-170;
Pediatric
prole:
0-140;Infant
prole:0-120
0-200
(Adolescent
prole:
0-170;
Pediatric
prole:
0-140;Infant
prole:
0-120)
HemoSweep
Speed
Selectthewaveform
displaysweepspeed.
25mm/s
DisplayFormatSelectthepressure
valuesdisplayformat.
Sys/Dia(Mean)
PrimaryHRSource
Selecttheparameterto
calculateHR.
AUTO
(USonly:ECG)
ShowPulseRateEnablethepulserate
display.
Disabled
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>Art1>Advancedtoadjust
thesesettingsforthecurrentpatient.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUORPACU
FilterHzSelectthewaveform
lter.
12
Catheter
Disconnect
Enablethedisconnect
alarm.
Enabled
360CARESCAPEONE2094480-001