SpO₂
SettingtheSatSeconds
™
alarmmanagementthreshold
NOTE
Nellcor
™
sensorswithOxiMax
™
technologyonly.
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page3.
3.SetthethresholdwiththeSaturationSecondsarrows.
SettingtheSpO
2
alarmsandalarmlimits
1.SelecttheSpO
2
parameterwindow.
2.SelecttheAlarmstab.
3.SetthealarmlimitsforSpO2,HR,orPRSpO2.
Ifafeatureisnotactive,alarmlimitsaregreyedout.SelectAlarmOntosetthe
alarmlimits.HRappearswhentheHRAlarmsoptionissettoSingleintheCare
UnitSettings.PRSpO2appearswhentheHRAlarmsoptionissettoMultiplein
theCareUnitSettings.Careunitsettingsarepasswordprotected.Formore
information,seethesupplementinformationprovided.
4.Setthealarmlimits.
DeactivatingtheSpO2probeoffalarm
ThisfeatureismeanttobeusedwhenendingSpO
2
monitoring.Itshouldnotbeused
duringactiveSpO
2
monitoring.ThissettingcanbeenabledinDefaultSetupandthe
settingispasswordprotected.TherearetwooptionstodeactivatetheSpO
2
probeoff
alarm,fromtheSpO
2
menuorusingtheaudiopausekey.Ifithasbeenenabled,there
willbeabuttonontheSpO2tabthatallowsyoutodeactivatethealarm:
1.SelecttheSpO
2
parameterwindow.
2.SelectSetup>Page1.
3.SelectDeactivateSpO2ProbeOff.
Ifdeactivationwiththeaudiopausekeyhasbeenenabled,SpO2probeoff
deactivationcanbeacknowledgedbyselecting
twice.
Whenthealarmisdeactivated,therewillbenoaudibleorvisualSpO2probeoffalarm
indications.ThealarmisautomaticallyreactivatedifSpO
2
vitalsignsaredetectedand
alarmconditionismetagain.
WARNING
WithdeactivatedSpO2probeoffalarm,keepthepatient
underclosesurveillance.
SpO
2
alarmpriorities
YoucanselectprioritiesfortheSpO2high,SpO2low,andSpO2probeoffalarms
throughAlarmsSetup>AlarmPriorities>OtherParameters.Thechoicesare
Low,Medium,High,andEscalating.ForSpO2highandSpO2low,Lowpriorityis
onlyavailableifithasbeenenabledduringcongurationintheCareUnitSettings.
ForSpO2probeoff,LowandMediumprioritiesareonlyavailableiftheyhave
beenenabledduringcongurationintheCareUnitSettings.Careunitsettings
arepasswordprotected.Formoreinformation,seethesupplementalinformation
provided.
148CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017