Invasivepressures
■Wheneverthepatient’spositionischanged.
■Wheneverthepressurereadingisquestionable.
●PressurescanbezeroedindividuallybyselectingZerointhepressuremenu.You
canzeroallpressuresexceptICPbyselectingZeroAllPressuresinthemainmenu.
Zeroingtheinvasivepressuretransducers
1.Levelthetransducerfollowingyourcareunit’spolicy(usuallylevelofthe
phlebostaticaxis).
2.Closethetransducerstopcocktothepatientandopentheventingstopcocktoair.
3.Ifthepressurelineyouaretryingtozerodoesnothavethetransduceropentoair,
themessagePressureSenseddisplays.
4.Youcanzeroallconnectedpressuretransducerssimultaneouslybyselecting
ZeroAllPressuresfromthemainmenu,oryoucanzeroasingleactivepressure
transducerbyselectingtheinvasivepressureparameterwindow>Setup>Page
1>Zero.
NOTE
ZeroAllPressuresdoesnotzeroaconnectedICPchannel.
TheICPchannelmustbezeroedseparately.Whenthe
ZeroICPseparatelymessagedisplays,youcanzerothe
ICPchannelbyselectingZerofromtheICPSetup>Page
1tab.
5.Checkthatazeroreferencehasbeenestablished.Watchthepressureparameter
windowformessages.
6.Closetheventingstopcocktoairandopenthetransducerstopcocktothepatient.
7.Checkthatpressurenumericsdisplayonscreen.
Selectinganinvasivepressurechannellabel
1.Selecttheinvasivepressureparameterwindow.
2.SelectSetup>Page1.
3.SelectachannellabelfromtheLabellist.
ThechannellabelPAcanonlybeusedforonechannellabel.Allotherlabels
namescanbeusedformorethanonechannellabel,ifdesired.
Selectingthesizeoftheinvasivepressurewaveform
1.Selecttheinvasivepressureparameterwindow.
2.SelectSetup>Page1.
3.SetthewaveformscalewiththeScalearrows.
Thelargerthescalevalue,thesmallerthewaveformsize.
Optimizingtheinvasivepressurewaveformscale
Youcanselectanautomaticcalculationforanoptimizedwaveformsize.Thissizewill
thenbeusedforthewaveform.