Alarms
3.SelectAutoLimits.
Ifyouneedtoundothesechangesandreturntothepreviousalarmlimitsettings,
selectUndoSettingsbeforeclosingthemenu.
Defaultautoalarmlimits
Parameter
HighlimitLowlimit
NIBP
Adultandchildcuff:S/D/M:NIBP*1.25+10
Infantcuff:S/D/M:NIBP*1.25+2
Adultandchildcuff:S/D/M:NIBP*0.75-10
Infantcuff:S/D/M:NIBP*0.75–2
AllHR/PRparameters
(ECG,SpO
2
,UAC,Art,
Fem)
AllHR*1.25ofthecurrentHRvalueAllHR*0.75ofthecurrentHRvalue
PVCPVC+10
Notapplicable
SVCSVC+10
Notapplicable
EtCO
2
EtCO
2
+1%EtCO
2
-1%
SpO
2
SpO
2
+5%SpO
2
-5%
Art,Fem
Sys/Dia/Mean:
Value*1.25+10mmHg
Value*1.25+1.3kPa
Sys/Dia/Mean:
Value*0.75-10mmHg
Value*0.75-1.3kPa
FemV,CVP,PA,RAP,
RVP,LAP,ICP,CPP,P1,
P2
Mean:
Value*1.25+5mmHg
Value*1.25+0.67kPa
Mean:
Value*0.75-5mmHg
Value*0.75-0.67kPa
UAC
Sys/Dia/Mean:
Value*1.25+5mmHg
Value*1.25+0.67kPa
Sys/Dia/Mean:
Value*0.75-5mmHg
Value*0.75-0.67kPa
UVC
Mean:
Value*1.25+5mmHg
Value*1.25+0.67kPa
Mean:
Value*0.75-5mmHg
Value*0.75-0.67kPa
Temperature
Tx+1°C
Tx+1.8°F
Tx-1°C
Tx-1.8°F
Tx-Ty(e.g.,T2-T1)
Tx-Ty+1°C
Tx-Ty+1.8°F
Tx-Ty-1°C
Tx-Ty-1.8°F
RR
RR*1.25+2RR*0.75-2
Returningthedefaultalarmlimits
1.SelectAlarmsSetupfromthemainmenu.
2.SelecttheAlarmLimitstab.
3.SelectDefaultLimits.
Ifyouneedtoundothesechanges,selectUndoSettingsbeforeclosingthemenu.
Beawarethatreturningtothedefaultalarmlimitswillalsoreturnthestoredalarm
on/offstatus,andusingtheUndoSettingsoptionswillnotrevertanychangesto
thealarmon/offstatus.
80CARESCAPEONE2094480-001