Alarmspecications
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
Temperature
malfunction
alarm
0s
Tempdevice
overheated.
Shuttingdown
alarm
0s
AlarmprioritiesandescalationtimesforCO
2
Thefollowingtableshowsthemessageswiththeirprioritiesandescalationtimes.Inthe
table,themessagelocationisindicatedbythefollowing:alarm=alarmarea;param.=
parameterwindow;andwavef.=waveformarea.Inthefollowingtable,0s=nointended
delay.Formoreinformationonalarmmessages,refertotheuserinformationprovided.
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
Apnea(CO2)
ORswpackage,
adjustablelimit
seconds
alarm,
wavef.,
param.
0s
Xseconds
(X=
adjustable
limit)
X+40sX+102s
Apnea(CO2)
PACU,ED,ICU,
NICUsw
packages,
adjustablelimit
seconds
alarm,
wavef.,
param.
0s
Xseconds
(X=
adjustable
limit)
X+20sX+40s
Apnea(CO2)
Withalarm
prioritysetto
mediumorhigh
alarmXseconds(X=adjustable
limit),accordingtopriority
setting:high,medium
Apnea
deactivated
param.
0s
CheckAdapter
param.
0s
CheckCO2
adapter
alarm
40s60s
Checkdevice
param.
0s
Checksample
line
alarm,
param.
0s40s60s
Checksensor
param.
0s
CO2device
failure:Call
service
alarm
0s
CO2device
initializing
alarm
0s
CO2device
malfunction
alarm
0s
CO2device
overheated
Shuttingdown
alarm
0s
292CARESCAPEONE2094480-001