EasyManuals Logo

GE CARESCAPE ONE User Manual

GE CARESCAPE ONE
394 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #226 background imageLoading...
Page #226 background image
Messages
Message
Location
PossiblecausesSuggestedactions
Tempdevicefailure:
Callservice
al.area
CARESCAPETemperaturestartup
orinitializationfailure.
ReconnecttheCARESCAPE
Temperature.
ReplacetheCARESCAPE
Temperature.
Iftheproblempersists,
contactqualiedservice
personnel.
Tempdevice
overheated.Shutting
down.
al.areaCARESCAPEONEturnsoff
theCARESCAPETemperature
becauseitssurfacetemperature
istoohigh.
ChecktheCARESCAPE
Temperatureplacementand
ensurethatitisnotcovered
byanythingthatmightlead
tooverheating.
ReplacetheCARESCAPE
Temperature.
MessagesrelatedtoCO
2
measurement
Forinformationregardingalarmprioritiesandescalationtimes,seethesupplemental
informationprovided.
Makesureyouarefamiliarwiththegenericlayoutofthescreen.Thiswillhelpyouidentify
whereonscreenthefollowingmessagesappear.Themessagelocationisindicatedwiththe
followingabbreviations:
al.area=alarmarea
param.=parameterwindow
wavef.=waveformarea
Message
Location
PossiblecausesSuggestedactions
Apnea(CO2)
APN
Apnea
al.area
param.
wavef.
Nobreathingdetected.Checkthepatientstatus.
Wait.Themessage
disappearsafterabreathis
detected.
Apneadeactivated
param.
Anewcasehasjustbeen
started/anewpatientadmitted,
orthemeasurementhasjust
beenstartedandtheapnea
alarmisnotactiveyet,orthe
userhasdeactivatedtheapnea
alarm.
Wait.Themessage
disappearsafter3breaths
aredetectedwithinone
minute.
226CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE CARESCAPE ONE and is the answer not in the manual?

GE CARESCAPE ONE Specifications

General IconGeneral
CategoryMedical Equipment
ManufacturerGE Healthcare
ModelCARESCAPE ONE
Display Size10.1 inches
ECGYes
SpO2Yes
NIBPYes
TemperatureYes
RespirationYes
ParametersECG, SpO2, NIBP, Temperature, Respiration
Device TypePatient Monitor
ConnectivityBluetooth
Operating Temperature0° to 40°C (32° to 104°F)
Storage Temperature-20°C to 60°C (-4°F to 140°F)
Humidity15 to 95% RH, non-condensing
Display TypeTouchscreen LCD

Related product manuals