Messages
Message
Location
PossiblecausesSuggestedactions
●CheckCO2adapter
●CheckAdapter
●al.area
●param.
Thereisanobstructionin
theLoFloreceptaclesample
cell,ortheLoFloreceptacle
samplecellisnotwithinthe
optimaloperatingrangeforCO
2
measurement.
●Ensurethatthesamplingline
connectoriscorrectlyand
fullyinserted.
●Checkthatthesamplecell
windowsandreceptacle
windowsdonothaveany
foreignmaterialonthem.
●Performzeroingprocedureto
optimizetheoperation.
●Iftheproblempersists,
replacetheadapter.
●Checksampleline●al.area
●param.
Sampleowrateisloweror
higherthanthenominalow
duetoanobstructioninthe
samplelineorexhaustport,ora
cutorsplitinthesampleline.
Thelterinthesamplelineis
completelysaturated.
Ifthesituationpersistsformore
than2minutes,thepumpwill
turnoff.
●Checkthesamplelineand
exhaustportforobstructions.
●Checkthesamplelinefor
cutsorsplitsandreplaceif
necessary.
●Removethesampleline
anddisposeofitfollowing
hospitalprotocol.
●Ifthepumphasshutoff,
disconnectthenreconnect
theaccessorytorestartthe
pump
●CO2devicefailure:
Callservice
●al.area
CARESCAPECO
2
startupor
initializationfailure.
●ReconnecttheCARESCAPE
CO
2
.
●ReplacetheCARESCAPECO
2
.
●Contactqualiedservice
personnel.
●CO2deviceinitializing●al.areaTheCARESCAPECO
2
isinthe
softwaredownloadmode.
●Waitandseeifthemessage
disappears.
●ReplacetheCARESCAPECO
2
.
●Contactqualiedservice
personnel.
●CO2device
malfunction
●Checkdevice
●al.area
●param.
CARESCAPECO
2
startupor
runtimefailure.
●ReconnecttheCARESCAPE
CO
2
.
●ReplacetheCARESCAPECO
2
.
●Contactqualiedservice
personnel.
●CO2device
overheatedShutting
down
●al.areaTheCARESCAPECO
2
surface
temperatureistoohighand
theCARESCAPEONEturnsthe
measurementoff.
●ChecktheCARESCAPECO
2
placementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
●ReplacetheCARESCAPECO
2
.
●Contactqualiedservice
personnel.