Messages
Message
Location
PossiblecausesSuggestedactions
●CO2measurement
removed
●al.areaTheCARESCAPECO
2
has
beendisconnectedfromthe
CARESCAPEONE.
●ConnecttheCARESCAPECO
2
andLoFloSidestreamModule
ifyouwanttorestartthe
measurement.
●CO2sensorremoved
●Sensorremoved
●al.area
●param.
TheLoFloSidestreamModule
hasbeendisconnectedfromthe
CARESCAPECO
2
.
●ConnecttheLoFloSidestream
ModuletotheCARESCAPE
CO
2
ifyouwanttorestartthe
measurement.
●CO2zerorequired
●Zerorequired
●al.area
●param.
TheCARESCAPECO
2
needs
zeroing.
●Disconnectthepatientand
performzeroingfromthe
CO
2
menu:Zeroing>Zeroto
roomair.
●Iftheproblempersists,
replacethesamplinglineand
zeroagain.
●EtCO2high
●EtCO2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Checktheventilatorand
freshgasowsettings.
●Checkarterialbloodgas
values.
●Adjustalarmlimitsif
necessary.
●FiCO2high
●FiCO2low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Checktheventilatoror
absorber,breathingcircuit,
andfreshgasowsettings.
●Adjustalarmlimitsif
necessary.
●IdenticalCO2device●al.areaThereismorethanone
CARESCAPECO
2
devicein
thesystem.
●Removeallbutone
CARESCAPECO
2
.
●MoveCO2devicetoa
coolerlocation
●Deviceoverheating
●al.area
●param.
TheCARESCAPECO
2
surface
temperatureistoohigh.
●ChecktheCARESCAPECO
2
placementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
●ReplacetheCARESCAPECO
2
.
●Contactqualiedservice
personnel.