Defaultsettings
Setting
ED
ICUNICUORPACU
CVPMeanHigh
Limit/LowLimit
15/0
(Infantprole:10/0)
5/015/0
(Infantprole:10/0;Perfusion
prole:15/-5)
CVPMeanAlarmOnEnabledEnabled
(Traumaand
Neuroproles:
Disabled)
Enabled
UVCMeanHigh
Limit/LowLimit
15/0
UVCMeanAlarmOnDisabled
FemVMeanHigh
Limit/LowLimit
15/0
(Infantproles:10/0)
5/015/0
(Infantproles:10/0)
FemVMeanAlarmOnEnabled
SelectPA,ICP&P1-P2tabtoadjusttheinvasivepressuresalarmprolesettings.The
followingtableliststhefactorydefaultsettings.
Setting
ED
ICUNICUORPACU
PASysHighLimit/Low
Limit
40/10
PASysAlarmOnDisabled
(USonly:Enabled)
PAMeanHigh
Limit/LowLimit
30/5
PAMeanAlarmOnEnabled
PADiaHighLimit/Low
Limit
20/5
PADiaAlarmOnEnabled
(Traumaand
Neuroproles:
Disabled)
Enabled
ICPMeanHigh
Limit/LowLimit
15/0
(Infantprole:9/0)
9/015/0
(Infantprole:9/0)
ICPMeanAlarmOnEnabled
CPPHighLimit/Low
Limit
100/60
(Infantprole:90/50)
100/40100/60
(Infantprole:90/50)
CPPAlarmOnEnabled
336CARESCAPEONE2094480-001