Parameterspecications
Pulseoximetrydisplaydataupdate
period
<30s
Pulseoximetrysweepspeedoptions
6.25,12.5,25,and50mm/s
PulseoximetrywaveformscaleoptionsTruSignaltechnology:AUTO,50,20,10,5,2
MasimoandNellcortechnologies:1x,2x,4x,and8x
WavelengthinformationforSpO
2
probe
LEDs
Thisinformationcanbeespeciallyusefultoclinicianssuchas
thoseperformingphotodynamictherapy.
GETruSignalpulseoximetrysensorLEDpeakwavelengthsare
within600to1000nmandthemaximumopticaloutputpowerfor
eachLEDislessthan15mW.
NellcorOxiMaxpulseoximetrysensorscontainLEDsthatemitred
lightatawavelengthofapproximately660nmandinfraredlight
atawavelengthofapproximately900nm.Thetotalopticaloutput
powerofthesensorLEDsislessthan15mW.
MasimoSETpulseoximetrysensorscontainLEDsthatemitred
lightatawavelengthofapproximately660nmandinfraredlight
atawavelengthofapproximately905nmforLNOPandLNCS,
approximately663nmand880nmforLNOPandLNCStipclips,
andapproximately660nmand880nmforLNOPandLNCS
transectanceforeheadsensors.Thetotalopticaloutputpower
oftheLEDsislessthan15nW.
PulseoximetryparametersmonitoredArterialoxygensaturation(SpO
2
)andpulserate
Pulseoximetrysaturationmeasurement
valueanddisplayrange
GETruSignal:0to100%
Masimo:0to100%
Nellcor:1to100%
Pulseoximetryperipheralpulserate
range
GETruSignal:30to300bpm
Masimo:25to240bpm
Nellcor:20to300bpm
Thespeciedaccuracyforeachtechnologyisthe
root-mean-square(RMS)differencebetweenthemeasuredvalues
andthereferencevalues.Becausepulseoximetryequipment
measurementsarestatisticallydistributed,onlyabouttwo-thirds
ofthepulseoximetryequipmentmeasurementscanbeexpected
tofallwithinthe±1A
rms
ofthevaluemeasuredbyaCO-oximeter.
Actualaccuracydependsonsensor.Pleaserefertothesensor
instructionsforuseformoredetailedinformation.
GETruSignalaccuracy:
Withoutmotion:
SpO
2
(70to100%):±2Adult,±2Pediatric,±3Neonatal
SpO
2
(<70%):Unspecied
Withmotion:
SpO
2
(70to100%):±3Adult,±3Pediatric,±3Neonatal
SpO
2
(<70%):Unspecied
Lowperfusion:
SpO
2
(70to100%):±2Adult,±2Pediatric,±3Neonatal
SpO
2
(<70%):Unspecied
Pulseoximetrysaturationmeasurement
valueaccuracy
2094480-001CARESCAPEONE305
CARESCAPE ONE User Manual DRAFT 26 October 2017