Parameterspecications
Masimoaccuracy:
Thefollowingaccuracyspecicationsrepresentonlythedevice’s
portionoftheintegratedMasimorainbowSETtechnology
performance.Theactualmeasurementperformanceand
accuracydependsontheaccessoryusedandcanbelimitedby
theaccessoryasspeciedinthesensor’sinstructionsforuse.
●Withoutmotion:
■SpO
2
(70to100%):±2%Adult,±3%Neonate
■SpO
2
(<70%):Unspecied
●Withmotion:
■SpO
2
(70to100%):±3%Adult,±3%Neonate
■SpO
2
(<70%):Unspecied
●Lowperfusion:
■SpO
2
(70to100%):±2%Adult,±3%Neonate
■SpO
2
(<70%):Unspecied
Nellcoraccuracy:
Saturationaccuracyvariesbysensortype.ContactMedtronic
forsensoraccuracyinformation.Adultspecicationsareshown
forOxiMaxMAXAandMAXNsensorswiththepulseoximeter.
NeonatespecicationsareshownforOxiMaxMAXNsensorswith
thepulseoximeter.
●Withorwithoutmotion:
■SpO
2
(70to100%):±2Adult,±2Neonatal
■SpO
2
(60to80%):±3Adult,±3Neonatal
■SpO
2
(<60%):Unspecied
●Lowperfusion:
■SpO
2
(70to100%):±2Adult,±2Neonatal
■SpO
2
(<70%):Unspecied
Pulseoximetryperipheralpulserate
accuracy
Thespeciedaccuracyforeachtechnologyisthe
root-mean-square(RMS)differencebetweenthemeasuredvalues
andthereferencevalues.Actualaccuracydependsonsensor.
Pleaserefertothesensorinstructionsforuseformoredetailed
information.
GETruSignalaccuracy:
●Withoutmotion(30to250bpm):±2Adult,±2Pediatric,±2
Neonatal
●Withmotion(30to250bpm):±5Adult,±5Pediatric,±5Neonatal
●Lowperfusion(30to250bpm):±3Adult,±3Pediatric,±3
Neonatal
Masimoaccuracy:
●Withoutmotion(25to240bpm):±3bpmAdult,Pediatric,
Infant,Neonate
●Withmotion(25to240bpm):±5bpmAdult,Pediatric,Infant,
Neonate
306CARESCAPEONE2094480-001