ECG
AlarmmessageArrhythmiaanalysisAlarmprioritydefaultArrhythmiadetection
criteria
MissingBeat
Full
Allsoftwarepackages
exceptNICU
Accordingtopriority
setting
ActualRRintervalmore
than1.8timesthe
averageRRinterval.
MultifocalPVCsFullAccordingtopriority
setting
Within16previousbeats
twoormorePVCswith
differentmorphologies
aredetected.
Pause
FullAccordingtopriority
setting
Couplinginterval
betweentwobeats
exceeds1to5seconds
(congurable).
RonT
FullAccordingtopriority
setting
IsolatedPVCisdetected
within100msofthe
peakoftheT-waveofthe
patient’spredominant
normalbeat.
SVTachyFullAccordingtopriority
setting
ArunofSVCsisdetected
witharunlengthofat
leastthesetSVTLength
andtheheartrateisat
leastthesetHRforSVT
/min.
VBradyFullAccordingtopriority
setting
RunofPVCsaredetected
witharunlengthofat
leastthreebeats.In
addition,atleasttwo
consecutiveRRintervals
intherunmusthavean
effectiveheartrateless
than50bpm.
VFib/VTachFull,LethalHighECGwaveformindicates
achaoticventricular
rhythm.
VTachFull,LethalAccordingtopriority
setting.
AlwaysHighifVTach
duration>30seconds,
andHR≥180bpmin
NICUor≥150bpmin
othersoftwarepackages,
andHRisovertheuser
adjustedHRhighlimit.
ArunofPVCsisdetected
witharunlengthofsix
beatsormoreandthe
effectiveHRexceedsthe
selectedHRforVTach
/minvalue.
VT>2
FullAccordingtopriority
setting
ArunofPVCsisdetected
witharunlengthofmore
thantwobeatsbutless
thanthenumberrequired
forVTach.Inadditionat
leasttwoconsecutiveRR
intervalsintherunmust
haveaneffectiveHRthat
2094480-001CARESCAPEONE119
CARESCAPE ONE User Manual DRAFT 26 October 2017

Table of Contents

Other manuals for GE CARESCAPE ONE