Messages
Message
Location
PossiblecausesSuggestedactions
●Noprobe
●NoSpO2probe
●NoadhesiveSpO2
probe
●param.
●al.area
Sensorisnotconnectedtothe
CARESCAPESpO
2
device.
Adhesivesensorisnot
connectedtotheCARESCAPE
SpO
2
device.
●Checktheconnection
betweenthesensorand
theCARESCAPESpO
2
device.
●PR(SpO2)high/
PR(SpO2)low
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Probeinitializing
●param.
CARESCAPESpO
2
–Masimo
hasdetectedthatthesensoris
initializing.
●SpO
2
monitoringwillbegin
afterinitializationiscomplete.
●Probeoff
●SpO2ProbeOff
●param.
●al.area
Thesensormaybedefectiveorit
isnotconnectedtothepatient.
●Checkthepatientstatus.
●RepositiontheSpO
2
sensor.
●ReplacetheSpO
2
sensor.
●Pulsesearch
●param.
Themeasurementisstarting
normally.
Defectiveordamagedsensoror
cable.
Sensorisoffofthepatient.
Detectionofarepeatablepulse
hasstopped.
Pulseisnotrecognizedby
thealgorithmduetoalow
amplitudesignal.
●Waitforwaveformsand
parametervaluestodisplay.
●Checkthesensorandcable.
●Repositionorreplacesensor.
●Replacecable
●ReplaceSpO2cable
●param.
●al.area
CARESCAPESpO
2
–Masimohas
detectedthattheinterconnect
cablehasexpired.
●Replacetheinterconnect
cable.
●Replaceprobe
●ReplaceSpO2probe
●Replaceadhesive
SpO2probe
●param.
●al.area
CARESCAPESpO
2
–Masimohas
detectedthatthesensorhas
expired.
CARESCAPESpO
2
–Masimo
hasdetectedthattheadhesive
sensorhasexpired.
●Replacethesensor.
●SpO2devicefailure:
Callservice
●al.areaCARESCAPESpO
2
devicestartup
orinitializationfailure.
●ReconnecttheCARESCAPE
SpO
2
.
●ReplacetheCARESCAPESpO
2
.
●Iftheproblempersists,
contactqualiedservice
personnel.
●SpO2device
overheatedShutting
down
●al.areaThesurfacetemperatureofthe
CARESCAPESpO
2
istoohighand
theCARESCAPEONEturnsthe
CARESCAPESpO
2
off.
●ChecktheCARESCAPESpO
2
deviceplacementandensure
thatitisnotcoveredby
anythingthatmightleadto
overheating.
●ReplacetheCARESCAPESpO
2
device.